PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
37/107
Set and Query Saved Items of Normal Measurement Values in Respective Channels
Syntax
Command
:DATAout:ITEM:NORMal <U Data>,<I Data>,<P Data>,<S Data>,
<Q Data>,<PF Data>,<DEG Data>,<FREQ Data>
Query
:DATAout:ITEM:NORMal?
Response
<U Data>,<I Data>,<P Data>,<S Data>,<Q Data>,
<PF Data>,<DEG Data>,<FREQ Data>
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
U Data
-
-
-
-
U4
U3
U2
U1
I Data
-
-
-
-
I4
I3
I2
I1
P Data
-
-
-
-
P4
P3
P2
P1
S Data
-
-
-
-
S4
S3
S2
S1
Q Data
-
-
-
-
Q4
Q3
Q2
Q1
PF Data
-
-
-
-
PF4
PF3
PF2
PF1
DEG Data
-
-
-
-
DEG4
DEG3
DEG2
DEG1
FREQ Data
-
-
-
-
FREQ4 FREQ3 FREQ2 FREQ1
Description
Command
Sets the saved normal measurement value items to the
numerical value 0 - 255.
Query
Returns the saved normal measurement value items of the respective
channels as numerical values.
Example
Command
:DATA:ITEM:NORM 1,1,1,0,0,1,0,0
Set the voltage, current, effective power and power factor data of Channel 1
to ON.
Query
:DATA:ITEM:NORM?
Response
:DATAOUT:ITEM:NORMAL 1,1,1,0,0,1,0,0 (WHEN HEADER IS ON)
1,1,1,0,0,1,0,0 (WHEN HEADER IS OFF)
Note
Under this command, when U is turned to ON, all the U items in the specified channel
(:DATAout:ITEM:U) become ON, and when U is turned OFF, all the U items in the specified
channel become OFF.
(:DATAout:ITEM:I) become ON, and when I is turned OFF, all the I items in the specified
channel become OFF.
To
specify
individually
saved
items
in
the
U
data
and
I
data,
use :DATAout:ITEM:U, :DATAout:ITEM:I after this command is executed.
During Query, if even 1 item in the channel is turned to ON, the channel will become ON.