![Hioki PW3390 Instruction Manual Download Page 106](http://html.mh-extra.com/html/hioki/pw3390/pw3390_instruction-manual_2137353106.webp)
PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
106/107
About Status data
The status information shows the measurement status when the measurement data is being saved, and
is shown as a hexadecimal of 32bit.
Contents of the individual bits in the 32bit are shown below.
bit 31
bit 30
bit 29
bit 28
bit 27
bit 26
bit 25
bit 24
HM4
HM3
HM2
HM1
MRB
MRA
MPB
MPA
bit 23
bit 22
bit 21
bit 20
bit 19
bit 18
bit 17
bit 16
ULM
UDP
UCU
HUL
UL4
UL3
UL2
UL1
bit 15
bit 14
bit 13
bit 12
bit 11
bit 10
bit 9
bit 8
RI4
RI3
RI2
RI1
RU4
RU3
RU2
RU1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
PI4
PI3
PI2
PI1
PU4
PU3
PU2
PU1
HMx : Harmonic parameter invalid
(
when the Harmonic synchronization is off, etc.
)
MRx : Motor analysis function A, B range over
MPx : Motor analysis function A, B peak over
ULM : Motor analysis function A, B synchronization unlock
UDP : Undisplayable
(
when the measuremnt data becomes significantly invalid immediately
after a range change, etc.
)
UCU : Uncalculable
(
when the measuremnt data becomes invalid immediately after a range
change, etc.
)
HUL : Harmonic synchronization unlock
ULx : Respective channels synchronization unlock
RIx : Respective channels current range over
RUx : Respective channels voltage range over
PIx : Respective channels current peak over
PUx : Respective channels voltage peak over
(
x is the channel number
)
Summary of Contents for PW3390