PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
57/107
Set and Query Noise Lower Limit Frequency
Syntax
Command
:FFT:LOWerfreq <frequency data>
Query
:FFT:LOWerfreq?
Response
<frequency data>
:
<OFF/1kHz/2kHz/3kHz/4kHz/5kHz/6kHz/7kHz/8kHz/9kHz/10kHz>
Description
Command
Sets noise lower limit frequency.
Query
Returns the noise lower limit frequency setting as a character string data.
Example
Command
FFT:LOW 10kHz
Set the noise lower limit frequency to 10kHz.
Query
:FFT:LOW?
Response
:FFT:LOWERFREQ 10kHz (WHEN HEADER IS ON)
10kHz (WHEN HEADER IS OFF)
Note
Specify OFF when set to 0kHz.
The key upper limit may be restricted depending on the sampling speed setting.
Select and Query Noise Analysis Point Values
Syntax
Command
:FFT:POINt <Point value>
Query
:FFT:POINt?
Response
<Point value>
:
<1000/5000/10000/50000>
Description
Command
Sets the noise analysis point value.
Query
Returns the setting for the noise analysis point value as a numerical value.
Example
Command
:FFT:POIN 1000
Set the noise analysis point value to 1000.
Query
:FFT:POIN?
Response
:FFT:POINT 1000 (WHEN HEADER IS ON)
1000 (WHEN HEADER IS OFF)