PW3390 Power Analyzer Communication Command Instruction Manual Second Edition
80/107
Set and Query Voltage Data Communication Output Items
Syntax
Command
:MEASure:ITEM:U <CH 1>,< CH 2>,< CH 3>,< CH 4>,
<SUM12>,<SUM34>,<SUM123>
Query
:MEASure:ITEM:U?
Response
<CH 1>,< CH 2>,< CH 3>,< CH 4>,<SUM12>,<SUM34>,<SUM123>
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
CH 1
PK-
PK+
THD
FND
DC
AC
MN
RMS
CH 2
PK-
PK+
THD
FND
DC
AC
MN
RMS
CH 3
PK-
PK+
THD
FND
DC
AC
MN
RMS
CH 4
PK-
PK+
THD
FND
DC
AC
MN
RMS
SUM12
-
-
-
-
-
-
MN
RMS
SUM34
-
-
-
-
-
-
MN
RMS
SUM123
-
-
-
-
UNB
-
MN
RMS
Description
Command
Sets the voltage data communication output items as 0
– 255.
Query
Returns the voltage data communication output items as
numerical values.
Example
Command
:MEAS:ITEM:U 1,1,1,1,0,0,0
Set the voltage RMS of Channels 1
– 4 to ON.
Query
:MEAS:ITEM:U?
Response
:MEASURE:ITEM:U 1,1,1,1,0,0,0 (WHEN HEADER IS ON)
1,1,1,1,0,0,0 (WHEN HEADER IS OFF)
Note
Use this command after :MEASure:ITEM:NORMal or :MEASure:ITEM:SUMCommand.
Summary of Contents for PW3390