82
Table 7.3 Setting Control Program Messages 8/10
Mnemonic Description
Format
S E Q
Sequential
mode
setting
d1 (Mode: 0 ~ 1)
0: OFF 1: ON
d2 (Program No.: 0 ~ 9)
No. of program executed when sequential mode
is turned ON
d3 (Discharging time before measurement)
(0.0 ~ 999.9s)
d4 (Charging time)
(0.0 ~ 999.9s)
d5 (Measuring time)
(0.0 ~ 999.9s)
d6 (Discharging time after measurement)
(0.0 ~ 999.9s)
[Note] Sequential mode: Subsequent parameters
are effective even on OFF.
(Retained as current set value)
[Format]
SEQ d1
d2, d3, d4, d5, d6
d1: NR1 format
d2: NR1 format
d3: NR2 format
d4: NR2 format
d5: NR2 format
d6: NR2 format
S E Q ?
Same as set items of query SEQ command
message in sequential mode. Returns current
set data.
d1 (Mode: 0 ~ 1)
d2 (Program No.: 0 ~ 9)
d3 (Discharging time before measurement)
d4 (Charging time)
d5 (Measuring time)
d6 (Discharging time after measurement)
[Format]
SEQ?
[Response]
C N F
Operation environment setting
d1 (Interlock control provided, not provided: 0 ~
1)
0: CONNECT 1: CUTOFF
d2 (Beep tone provided, not provided: 0 ~ 1)
0: OFF 1: ON
d3 (Beep tone good/no good: 0 ~ 1)
0: NO 1: GO
d4 (Click tone provided, not provided: 0 ~ 1)
0: OFF 1: ON
[Format]
CNF d1, d2, d3, d4, d5
d1: NR1 format
d2: NR1 format
d3: NR1 format
d4: NR1 format
C N F ?
Query response of operation environment is
same as setting.
[Format]
CNF?
[Response]
d1, d2, d3, d4
Summary of Contents for DSM-8104
Page 2: ......
Page 20: ...10 3 2 Block Diagram...
Page 82: ...72 Syntax Chart Fig 7 1 Message Syntax Chart...
Page 83: ...73 Command program header Fig 7 2 Program Header Syntax Chart...
Page 84: ...74 NRf format Fig 7 3 Syntax Chart of Data Part...
Page 130: ...120 11 Product Full View...
Page 131: ......
Page 132: ......