118
and execute a command from an interface or execute self-calibration by keying the keys on
the panel periodically.
[Note]
When sequential measurement mode is set to ON, auto self-calibration function is not
performed even if the self-calibration function is set to ON. When performing sequential
measurement, periodically perform self-calibration by panel operation or by an interface
command.
10.4 Self-diagnosis
Function
The self-diagnosis function consists of a function, by which self-diagnosis is executed by a
command by keying keys on the panel or by a command from an interface, and a function,
by which internal operations are always monitored in a normal operating condition.
See Paragraph 6.7.3 “Self-diagnosis” for the self-diagnosis function executed by keying.
In case malfunction is detected by a function that monitors internal operations, the meter
stops its operation and displays the following error message screen.
In this case, a component of the meter is faulty. Please contact the HIOKI sales office
near your location.
Error message:
“ERROR: xxx Call Service Center”
“xxx” in error codes have the following meanings.
Error Code
Description
001
Data in the memory is destroyed
002
Memory write error.
003
Memory read/write error.
004
Poor calibration of current measuring unit.
005
Poor calibration of A/D converter.
006
Internal communication error with measuring
unit.
007
Internal communication error with voltage
output unit.
008
A/D data readout error.
009
Power source frequency detection error.
010
Malfunction of voltage output unit.
[Note]
If meter operation is stopped by an error, pressing the [F3] SAVE key and
then the “- MAN.T” key sometimes resets meter operation to a normal operation
condition in about 60s.
Switching the power on again may return it to normal, but if an error persists, please
contact the nearest HIOKI office for repair.
Summary of Contents for DSM-8104
Page 2: ......
Page 20: ...10 3 2 Block Diagram...
Page 82: ...72 Syntax Chart Fig 7 1 Message Syntax Chart...
Page 83: ...73 Command program header Fig 7 2 Program Header Syntax Chart...
Page 84: ...74 NRf format Fig 7 3 Syntax Chart of Data Part...
Page 130: ...120 11 Product Full View...
Page 131: ......
Page 132: ......