![Hioki DSM-8104 Instruction Manual Download Page 36](http://html1.mh-extra.com/html/hioki/dsm-8104/dsm-8104_instruction-manual_2137277036.webp)
26
Table 5.4 Measuring Conditions
Set Item
Description
Setting Range (Specified Value)
*3
Resolution
MES.V Set
voltage
range
0.1 ~ 250 V
251 ~ 1000 V
(0.1,0.5,1.0,2.5,5.0,10,25,
50,100,250,500,1000)
0.1 V
1 V
[ F2 ] DOWN
[ F3 ] UP
SAMPL Integral time
Time setting 2 ~ 300 ms
(2,4,8,16,20,40,80,160,300)
Period setting 1 ~ 15 PLC
* 1
(1,2,4,8,15)
1 ms
1 PLC
[ F2 ] DOWN
[ F3 ] UP
[ F4 ] ms
[ F5 ] PLC
DELAY Trigger
delay
time
0 ~ 9999 ms
(0,5,10,50,100,500,1000,
5000,9999
)
1 ms
[ F2 ] DOWN
[ F3 ] UP
AVE
Averaging
ON/OFF
[ F4 ] ON
[ F5 ] OFF
Range 1 ~ 8 AUTO/HOLD
Range Integral Capacity
1 10
µF
2 1
µF
3 100
nF
4 10
nF
5 1
nF
6 100
pF
7 10
pF
8 10
pF
*2
RANGE Current
range
[ F2 ] DOWN
[ F3 ] UP
[ F4 ] AUTO
[ F5 ] HOLD
SEQ Sequential
measurement
Program No. 0 ~ 9
ON/OFF
[ F2 ] DOWN
[ F3 ] UP
[ F4 ] ON
[ F5 ] OFF
DCHG1 Discharging
time
(Before
measurement)
0 ~ 999.9 s
(0,10,20,30,40,50,60,600,900)
0.1 s
[ F2 ] DOWN
[ F3 ] UP
CHARG Charging time 0 ~ 999.9 s
(0,10,20,30,40,50,60,600,900)
0.1 s
[ F2 ] DOWN
[ F3 ] UP
MEAS Measuring
time
0 ~ 999.9 s
(0,10,20,30,40,50,60,600,900)
0.1 s
[ F2 ] DOWN
[ F3 ] UP
DCHG2 Discharging
time
(After
measurement)
0 ~ 999.9 s
(0,10,20,30,40,50,60,600,900)
0.1 s
[ F2 ] DOWN
[ F3 ] UP
NOTES
*1: “1 PLC” stands for “1 power line cycle.”
*2: “Range 8” increases the gain ten-fold while maintaining the same capacitor capacity
of 10pF.
*3: The specified value is a preset value selected by the scroll keys.
Summary of Contents for DSM-8104
Page 2: ......
Page 20: ...10 3 2 Block Diagram...
Page 82: ...72 Syntax Chart Fig 7 1 Message Syntax Chart...
Page 83: ...73 Command program header Fig 7 2 Program Header Syntax Chart...
Page 84: ...74 NRf format Fig 7 3 Syntax Chart of Data Part...
Page 130: ...120 11 Product Full View...
Page 131: ......
Page 132: ......