117
(1) Set the meter as follows.
●
Integral time (SAMPL)
300ms
●
Trigger mode (TRIG)
Internal (INT)
●
Averaging (AVE)
ON
●
Self-calibration (CALIBRATION)
Auto execute (AUTO MODE: ON)
(INTERVAL) 60s
●
Measuring voltage (MES. V)
10V
●
Measuring mode (MODE)
Current measurement
(2) Connect reference resistance of 100k
Ω
between the “INPUT” and “OUTPUT”
terminals.
(3) Connect a voltmeter between the “OUTPUT” terminal and “GUARD” terminal.
(4) Set the measuring voltage to 10V.
(5) Set the measuring mode to the Current Measuring mode.
(6) Set the current measuring range to the 100
μ
A range (“HOLD 1”).
(7) Measure and calibrate the 100
μ
A range.
(8) Change the reference resistance to 1M
Ω
, set the current measuring range to the
10
μ
A range (“HOLD 2”) and similarly calibrate the 10
μ
A range.
(9) Calibrate each range by the same method by changing the reference resistance
and range. Calibrate the maximum sensitivity, 10pA range, at reference
resistance of 100G
Ω
and measuring voltage of 1V.
The relationship between the reference resistance and current measuring range is shown
below.
Measuring
Voltage
Reference
Resistance
Current Range
10V
100kΩ
100μA(「
HOLD1
」)
10V
1MΩ
10μA(「
HOLD2
」)
10V
10MΩ
1μA(「
HOLD3
」)
10V
100MΩ
100nA(「
HOLD4
」)
10V
1GΩ
10nA(「
HOLD5
」)
10V
10GΩ
1nA(「
HOLD6
」)
10V
100GΩ
100pA(「
HOLD7
」)
1V
100GΩ
10pA(「
HOLD8
」)
[Caution]
Note that the full-scale value of the current measuring range varies depending on
the integral time. (See 5.1.5 “Setting measuring range”)
10.3 Self-calibration
Function
This meter has a self-calibration function and can continue self-calibration at a preset
interval by setting an interval. By setting this function ON, an accuracy can be maintained
for a long time.
During a preshipment factory inspection, this meter is set to self-calibration function ON and
60s as a self-calibration interval.
Self-calibration is performed by measuring an offset current and gain of the current
measuring unit and measured results are corrected by providing reference voltage and
reference resistance inside.
Self-calibration requires about 2s to complete and this time sometimes presents a problem
in automatic measurement. If such problem occurs, set the self-calibration function OFF
Summary of Contents for DSM-8104
Page 2: ......
Page 20: ...10 3 2 Block Diagram...
Page 82: ...72 Syntax Chart Fig 7 1 Message Syntax Chart...
Page 83: ...73 Command program header Fig 7 2 Program Header Syntax Chart...
Page 84: ...74 NRf format Fig 7 3 Syntax Chart of Data Part...
Page 130: ...120 11 Product Full View...
Page 131: ......
Page 132: ......