99
Table 7.7 Error Register
Bit No. and Name
Event/status indicated when bit is true
bit7:
Not
used
bit6:MLE
Message length error
Set when the length of a message exceeds a
permissible range.
Reset by reading this register.
bit5:HDE
Message header error
Set when a message header that cannot be
recognized is received.
Reset after reading this register.
bit4:DFE
Data format error
The number of parameters has exceeded a
specified number.
There are parameters that cannot be recognized.
Reset by reading this register.
bit3:DRE
Data range error
Set when a parameter value is outside specified
range.
Reset by reading this register.
bit2:CNE
A cannot-execute command
Set when a command that cannot be executed is
received.
Reset by reading this register.
bit1:ISE
Internal
communication
error
Set when an internal communication error is found.
Reset by reading this register.
bit0:BDE
Environmental data destruction
Set when data of a backup RAM is destroyed.
Reset by reading this register.