![Hioki DSM-8104 Instruction Manual Download Page 72](http://html1.mh-extra.com/html/hioki/dsm-8104/dsm-8104_instruction-manual_2137277072.webp)
62
Setting Screen
Measured
value display
screen
[ELEC]
Setting planar
sample electrode
[ENTER]
Planar sample
measuring mode
[SIZE]
[ACTL]
(D1,D2,t)
[ENTER]
Setting liquid
sample electrode
Fig. 6.11 Concept of Electrode Constant Setting
6.5 Current
Measurement
Set the measuring mode to “Current Measurement” referring to 5.1.2 “Setting measuring
mode” when measuring a current.
(1) Connection
When the measuring power source of the meter is used, connect measuring
object such as a part and circuit between the “OUTPUT” terminal and “INPUT”
terminal on the panel of the meter as in measurement of insulation resistance.
The polarities of the “OUTPUT” terminal and “INPUT” terminal are “+” and “-”
respectively. Connect the terminals matching the polarities of the work. Be
sue to use a shielded-conductor cable as the measuring lead connected to the
“INPUT” terminal to prevent induced troubles such as noise and connect the
shielded conductor to the “GUARD” terminal.
Connect the “+” side of the work to the “INPUT” terminal and “-” side to the
“GUARD” terminal when connecting to a work that generates a current.
In this case also, be sure to use a shielded-conductor cable as the
measuring lead connected to the “INPUT” terminal and connect the shielded
conductor to the “GUARD” terminal.
6.6 Ending
Measurement
Press the [STOP] key when finishing measurement.
Pressing the [STOP] key lights up the [STOP] lamp beside the key. The measuring
voltage output will become 0V and trigger input will be prohibited in a [STOP] state.
The circuit between the “OUTPUT” terminal and “INPUT” terminal will become a
discharge state (discharge state with a current limiter connected). The state of the
“CHARGE” terminal will also become the same state.
Summary of Contents for DSM-8104