Hill-Rom® P500 Therapy Surface User Manual (158251 REV 3)
41
Electromagnetic Immunity Guidance
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The P500 is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of the P500 should make sure it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic
Environment—
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6 kV Contact
± 8 kV Air
± 6 kV Contact
± 8 kV Air
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the rel-
ative humidity should be
at least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
± 2 kV for Power
Supply Lines
± 1 kV for Input/
Output Lines
± 2 kV for Power
Supply Lines
± 1 kV for Input/
Output Lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV Line(s) to
Line(s)
± 2 kV Line(s) to
Earth
± 1 kV Line(s) to
Line(s)
± 2 kV Line(s) to
Earth
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage Dips, Short
Interruptions, and
Voltage Variations
On Power Supply
Input Lines
IEC 61000-4-11
< 5% U
T
(> 95% dip in U
T
)
for 0.5 cycles
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
< 5% U
T
(> 95% dip in U
T
)
for 5 s
< 5% U
T
(> 95% dip in U
T
)
for 0.5 cycles
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
< 5% U
T
(> 95% dip in U
T
)
for 5 seconds
Mains power quality
should be of a typical
commercial or hospital
environment. If the user
of the P500 requires con-
tinued operation during
power mains interrup-
tion, it is recommended
that the P500 be powered
from an uninterruptedly
power supply or a battery.
Power Frequency
(50/60Hz) Mag-
netic Field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency mag-
netic fields should be at
levels characteristic of a
typical location in a typi-
cal commercial or hospi-
tal environment.
Note: U
T
is the AC mains voltage prior to application of the test level.