8
Hill-Rom® P500 Therapy Surface User Manual (158251 REV 3)
5.
Wipe down the surface with chlorine bleach (50 ppm to 150 ppm)
or mild detergent and warm water followed by an approved
intermediate level disinfectant.
NOTE:
2.5 oz of bleach per 10 gal of water is approximately 100 ppm of
available chlorine.
NOTE:
If you turn the surface to clean it, make sure the cleaning solution does
not pool or flow on to the other side or edges of the surface. This may
permit fluid to get into the surface air outlets and zipper closures that
ordinarily are protected by the ticking flaps.
6.
Let the bleach or disinfectant remain in contact with the surface as
instructed in the manufacturer’s instructions.
7.
Remove the bleach or disinfectant, and rinse with warm water.
8.
Let the surface and/or coverlet completely air dry.
9.
If the surface was removed, install it on the bed. Go to “Install the
P500” on page 25.
Machine Wash the AMT Coverlet
1.
Remove the coverlet.
2.
Machine-wash the coverlet with chlorine bleach (50 ppm to 150
ppm) or detergent and an effective intermediate level disinfectant.
(For customers in the US, the disinfectant should be registered with
the Environmental Protection Agency.)
NOTE:
2.5 oz of bleach per 10 gal of water is approximately 100 ppm of
available chlorine.
•
Use the bleach or disinfectant as instructed in the
manufacturer’s instructions.
•
To determine the amount of bleach or disinfectant to use,
determine the amount of water in the washer, and follow the
manufacturer’s dilution instructions.
•
During the wash cycle, soak the coverlet in the disinfectant or
bleach.
•
Let the coverlet rinse thoroughly in clean water.