![Hill-Rom P005723 User Manual Download Page 28](http://html1.mh-extra.com/html/hill-rom/p005723/p005723_user-manual_2134189028.webp)
22
Hill-Rom® P500 Therapy Surface User Manual (158251 REV 3)
On
1.
Make sure the occupant is centered on the surface.
2.
Make sure the head of the bed is below 30° and the
Head Alarm
is off.
3.
Make sure the bed siderails are raised and locked on the
side that the occupant will turn toward.
4.
Press the
Confirm Rails Up
control until its indicator
stays on.
NOTE:
You must press the Confirm Rails Up control within two
minutes before or after you press the Turn Assist control.
5.
Press the applicable
Turn Assist R
or
L
control until its indicator comes on.
When the Turn Assist mode is on, the surface turns the occupant to the
applicable side, beeps one time, and then holds the occupant in the
turned position. During the turn period, help support the occupant and
do the necessary procedure.
At the end of the turn period, the unit beeps three times, the occupant is
returned to the flat position, and then the surface returns to its previous
mode.
Sleep
The Sleep Mode is used to temporarily turn off the air system to permit
occupants who are sensitive to sleep surface movement to sleep. The air
pressure in the surface is monitored, but the air pump does not operate
unless the air pressures fall below or raise above a preset level. After
eight hours, the surface will go into Normal mode.
R