Hill-Rom® P500 Therapy Surface User Manual (158251 REV 3)
7
NOTE:
Clean and disinfect surfaces with nano Ag+® Technology as you would
other surfaces. This product is not intended to protect users or others
against bacteria, viruses, germs, or other disease organisms. Always
disinfect Hill-Rom surfaces thoroughly after each use according to
facility/healthcare protocol. Refer to “Clean and Disinfect the P500 and
Advanced Microclimate™ Technology (AMT) Coverlet” on page 7.
Clean
We recommend that you clean the unit with detergent and warm water.
Do not use excessive liquid or harsh cleansers.
Steam Clean
Do not use any steam cleaning device on the unit. Excessive moisture
can damage mechanisms in this unit.
Clean Difficult Spots
To remove difficult spots or stains, we recommend that you use standard
household cleansers and a soft bristle brush. To loosen heavy, dried-on
soil, you may first need to saturate the spot.
Disinfect
When there is visible soil and also between occupant use, we
recommend that you disinfect the unit with an EPA registered (US only),
tuberculocidal, disinfectant.
Clean and Disinfect the P500 and Advanced
Microclimate™ Technology (AMT) Coverlet
We recommend that you wipe down the P500 to clean it. In cases of
heavy soil, you can machine wash the coverlet. We also recommend that
you clean the surface every 60 days for occupants who are on the
surface longer than 60 days.
1.
Make sure the bed and control unit are unplugged.
2.
Remove all linens.
3.
If applicable, remove the surface from the bed.
4.
Examine the condition of the surface. If there are holes, tears, or
other signs of damage or deterioration of the ticking, replace the
surface.