Hill-Rom® P500 Therapy Surface User Manual (158251 REV 3)
6.
Transport the occupant on the bed as necessary. Refer to “Transport
Position and Stability” on page 4 and also the bed manufacturer’s
user documentation.
WARNING:
When transport is complete, connect the unit to an applicable
power source. Failure to do so could cause injury.
7.
When transport is complete, do as follows:
a.
Install the control unit and hose assembly on the footboard (see
“Install the P500” on page 25).
b.
Connect the hose assembly to the surface.
c.
Loss of Power
If the power to the unit is lost, this occurs:
• The control unit shuts down.
• The surface stays at its current mode—Normal or Max Inflate—for
approximately 3 hours.
NOTE:
If the unit is in the Turn Assist mode, the system will automatically go
into Normal mode.
When power returns to the system, the system does a self-test to make
sure of correct operation. When the system completes and passes the
self-test, the display will go through the set up (see “Set Up the System
for Occupant Placement” on page 27).