Hill-Rom® P500 Therapy Surface User Manual (158251 REV 3)
19
RemindMe™ Settings (Alarm)
The RemindMe™ Settings alarm can be turned on to remind
the caregiver to do care tasks. You can set the alarm to come
on from 30 minutes up to 4 hours.
On
1.
Press the control until its indicator stays on.
2.
Press and release the control as applicable to set the interval time
from 30 minutes to 4 hours.
When the alarm is set, the control’s indicator stays on and the alarm
state will occur at the end of the set time.
Example—
if you want the alarm to come on in 2 hours, press the
control four times. When 2 hours have passed, the surface will go into
an alarm state.
Off
—press the control until the indicator goes off.
Alarm Volume Settings
The alarms have three levels of volume: low, medium, and high. The
default volume level is high.
Change the volume level
—press and release the
Volume
control until you reach the applicable volume level as shown
on the display.
Silence
When the surface is in an alarm state, the Silence control
turns off the alarm sound and the flashing light bar for 30
minutes. After 30 minutes, if the problem has not been
corrected, the alarm sound and the flashing light bar will
come on again.
On
—press the control until its indicator stays on.
NOTE:
If you press the Silence control because the surface is in an
Out of Bed
alarm state, if the occupant returns to the surface, the system
automatically enables the alarm.