Hill-Rom® P500 Therapy Surface User Manual (158251 REV 3)
21
Max Inflate
Max Inflate mode causes the surface to become very firm. After the
surface has been in Max Inflate mode for 30 minutes, an alarm comes
on, the control’s indicator flashes, and the surface goes into Normal
mode.
NOTE:
The Max Inflate mode has a two-minute warning—an alarm comes on
and its indicator flashes—to let you know the therapy mode time is
almost over and the surface will soon go into the Normal mode.
On
—press the control until its indicator stays on.
Add 30 minutes to the time remaining
—press the
Max Inflate
control.
Off
—if you want to stop the Max Inflate mode before the 30 minutes
have passed, press the
Normal
control until its indicator comes on.
Turn Assist
WARNING:
Before you put the surface into the Turn
Assist mode, make sure the bed
siderails are raised and locked on the
side that the occupant is to turn toward.
Failure to do so could cause injury.
The Turn Assist mode helps you temporarily
turn the occupant to the left or right. When you turn this mode on, the
head and seat bladders adjust, and an internal air bladder inflates to start
the turning process. After the occupant reaches 25° to 35°, the surface
will pause for approximately 15 seconds. After the 15 seconds, an
audible indicator comes on, the bladder quickly deflates, and the surface
returns to its previous mode.
Turn Right Turn Left