Hill-Rom® P500 Therapy Surface User Manual (158251 REV 3)
27
d.
Turn the hanger hooks approximately 90° toward the bed
to lock them into the correct position.
•
When you connect the hose to the surface and the control unit,
you hear a click to let you know the hose assembly is correctly
connected.
•
You attach the hose hanger to
one of the hooks on the bottom
of the control unit to help keep
the hose off the floor.
WARNING:
Put the power cord in a position where
persons will not trip over it and it is
away from bed mechanisms. Failure
to do so could cause injury or
equipment damage.
•
The power cord is in a position
where persons will not trip over it and it is away from the bed
mechanisms.
•
All functions of the P500 operate correctly. Refer to “Control
Unit Features and Use” on page 14.
•
All functions of the bed frame operate correctly. Refer to the
bed manufacturer’s user documentation.
Set Up the System for Occupant Placement
1.
Make sure the control unit is correctly installed on the footboard
(see “Install the P500” on page 25).
WARNING:
If an error shows on the display and the light bar is
flashing
orange
, do not put the occupant on the surface.
Injury could
occur.
2.
Press the
Power
switch to
On
. The unit does a self test to make sure
it is operating correctly. One of these will occur:
•
If the unit detects an error, an alarm comes on, the light bar
changes to orange or flashing orange, and the display shows
the problem. Call Hill-Rom Technical Support.
If the light bar
is flashing orange, do not use the surface
.