Hill-Rom® P500 Therapy Surface User Manual (158251 REV 3)
29
NOTE:
The unit adjusts its pressure levels according to the entered weight and
the surface head angle; this is Normal mode.
•
To change the pressure levels of the surface, go to “Comfort
Adjust” on page 20.
•
To change the surface mode, go to “Surface Modes” on page
20.
Transfer the Occupant to or from the Unit
1.
Put the
bed
in the transfer position (refer to the bed manufacturer’s
user documentation).
2.
Put the
surface
into the Max Inflate mode (see “Max Inflate” on
3.
Transfer the occupant to or from the unit as necessary (refer to the
bed manufacturer’s user documentation).
Transport the Occupant on the Unit
1.
Press the
Power
switch to
Off
.
2.
Disconnect the control unit from its power source. The surface will
stay at its current pressure levels for approximately 3 hours.
NOTE:
No controls are available when the system is disconnected from AC
power.
3.
Close the display on the control unit.
4.
Put the
bed
in the transport position (refer to the bed
manufacturer’s user documentation).
5.
If necessary, do as follows to leave the control unit and hose
assembly in the room:
a.
Press the disconnect switch on
the hose assembly to disconnect
it from the surface.
b.
Remove the control unit from the
footboard.
c.
Put the control unit and hose
assembly in a safe place that is
away from traffic.
Disconnect Switch