Reverse Trendelenburg
CAUTION:
Bed adjustment into the Reverse Trendelenburg position can
cause the hose to be caught between the bed and the floor.
Always maintain good air hose management. Failure to do so can
cause equipment damage.
Some beds have the Reverse Trendelenburg position that lowers the foot
section close to the floor. Take care that the hose is not caught between
the bed frame and floor when the bed adjusts to this position.
Troubleshooting
WARNING:
Only authorized persons should service the P500. Service by
unauthorized persons could cause injury or equipment damage.
An Error Shows on the Display; the Light Bar is Solid
Orange or Flashing Orange
• If
an error shows on the display, and the light bar is
solid orange
, call
Hill-Rom Technical Support for assistance as soon as possible.
• If an error shows on the display, and the light bar is
flashing orange
,
remove the occupant from the unit, and then call Hill-Rom Technical
Support for assistance.
The Unit Does Not Go into Turn Assist Mode
• If you did not press the
Confirm Rails Up
control within
two minutes before or after you press the Turn Assist
control, Turn Assist mode will not operate.
• If the head of bed angle is more than 30°, Turn Assist
mode will not operate.
For either of the above, refer to the instructions for “Turn Assist” on
page 21.
An Alarm Does Not Fully Turn Off
If you press the
Silence
control, the alarm turns off
temporarily
. To