Hill-Rom® P500 Therapy Surface User Manual (158251 REV 3)
15
NOTE:
The CPR mode has a two-minute warning—an alarm comes on and the
light bar flashes—to let you know the CPR mode time is almost over
and the surface will soon go into the Normal mode.
On—
press the control until the light bar is red. The display will show
these:
and then
Add 60 minutes to the time remaining
—press the
CPR
control.
Off
—if you want to stop the CPR mode before the 60 minutes have
passed, press a surface mode control until its indicator comes on and the
light bar is green.
Light Bar
WARNING:
The occupant should not be on the
surface if the light bar is
flashing
orange
because of a system error.
Injury could occur.
When the control unit is connected to AC
power, the light bar on the front of the
unit shows the system’s status:
• Not on
—a weight has not been entered.
•
Green
—the system is operating correctly, and the system is not in
CPR mode.
•
Red
—the system is in CPR mode.
•
Orange
—the system is not in CPR mode, and an error has been
detected; the surface pressures
can
adjust.
•
Flashing orange
—the system is not in CPR mode, and the system is
in an alarm condition or an error has been detected. If an error has
been detected, the surface pressures
can not
adjust.