4.7 Stretcher Hydraulic Cylinder
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 26
DuraStar™ Series Stretcher Service Manual (man268)
16. Raise the stretcher to the high position.
17. If more than 24 pumps are required to raise the stretcher to the high
position, repeat the purge procedure in step 15 until the stretcher can be
raised to the high position with only 24 pumps or less.
18. Lower the stretcher to the low position, and verify that the cylinders are
descending at the same rate.
19. To ensure proper operation of the DuraStar™ Hydraulic Stretcher, perform
the “DuraStar™ Hydraulic Stretcher Function Checks” on page 2-4.
Summary of Contents for DuraStar Series
Page 57: ...Chapter 3 Theory of Operation Page 3 2 DuraStar Series Stretcher Service Manual man268 NOTES ...
Page 95: ...Chapter 5 Parts List Page 5 2 DuraStar Series Stretcher Service Manual man268 NOTES ...
Page 97: ...Warranty Chapter 5 Parts List Page 5 4 DuraStar Series Stretcher Service Manual man268 NOTES ...
Page 139: ...Chapter 6 General Procedures Page 6 2 DuraStar Series Stretcher Service Manual man268 NOTES ...