7.1 Patient Tray
Chapter 7: Accessories
Page 7 - 4
DuraStar™ Series Stretcher Service Manual (man268)
7.1
Patient Tray
Tools required:
None
The patient tray mounts on the top of the siderails when they are in the up and
locked position. When properly installed, the patient tray can support up to 45
lb (20 kg).
Installation
1. Put the siderails in the up and locked position (see figure 7-1 on page 7-4).
Figure 7-1. Patient Tray Installation
2. Slide the stationary hook (A) end of the tray onto one of the siderail top
rails.
3. Rotate the handle end (B) of the tray down toward the other siderail until
the handle snaps over the top of the siderail.
Removal
1. Pull up on the handle (B), releasing one side of the tray from the top of the
siderail.
2. Lift the tray off the siderails.
A
B
m268_025