7.4 IV Transporter
Chapter 7: Accessories
Page 7 - 8
DuraStar™ Series Stretcher Service Manual (man268)
7.4
IV Transporter
Tools required:
Drill
Ratchet
1/4" drill bit
Torque wrench
7/8" drill bit
3/4" socket
Retaining ring removal/installation tool
The IV transporter attaches to the base of the stretcher and enables the
transport of a portable IV pole without the assistance of the caregiver.
The IV transporter can be mounted on either the right or left side of the
stretcher at the head end only.
Installation
1. Decide on which side of the stretcher you will mount the transporter.
NOTE:
The proper location for the transporter hole is indicated by dimples located on
the inside surface on both sides of the head end of the shroud.
2. Lift the shroud, and locate the dimple at the head end of the shroud that
corresponds to the location where you want to mount the IV transporter.
3. Using the drill and the 1/4" drill bit, drill a 1/4" (6.4 mm) pilot hole through
the center of the dimple.
4. Using the drill with the 7/8" drill bit and the pilot hole as a guide, carefully
drill a 7/8" (22.4 mm) hole through the shroud.
5. Install the rubber grommet (D) into the hole you just drilled in the shroud
Summary of Contents for DuraStar Series
Page 57: ...Chapter 3 Theory of Operation Page 3 2 DuraStar Series Stretcher Service Manual man268 NOTES ...
Page 95: ...Chapter 5 Parts List Page 5 2 DuraStar Series Stretcher Service Manual man268 NOTES ...
Page 97: ...Warranty Chapter 5 Parts List Page 5 4 DuraStar Series Stretcher Service Manual man268 NOTES ...
Page 139: ...Chapter 6 General Procedures Page 6 2 DuraStar Series Stretcher Service Manual man268 NOTES ...