Understanding and addressing operational issues
8.6
Resetting the i.Center monitoring system
NOTE:
This section only applies to i.Series models.
Use the following procedure to reset the i.Center monitoring system.
To reset the monitoring system
1
From the battery backup, remove one battery.
2
Disconnect the incubator from power.
3
Connect the incubator to power. For more information, see Section 3.5.1, “Connecting to AC power.”
4
Replace the battery.
8.7
Viewing manufacturer and product information
NOTE:
This section only applies to i.Series models.
You can view software version information for control and display components of the i.Center monitoring
system. You can also view information for contacting Helmer.
To view manufacturer and product information
1
From the MAIN screen, use the buttons to navigate to and select the Product/Company Information
option. The Helmer logo appears briefly, followed by the INFORMATION screen. The INFORMATION
screen displays the Control Version and the Display Version.
2
Press the NEXT button. The manufacturer information appears.