S2M
TEDS TRANSDUCER IDENTIFICATION
24
9
TEDS TRANSDUCER IDENTIFICATION
TEDS (Transducer Electronic Data Sheet) allows you to store sensor characteristic values
in a chip as per IEEE 1451.4. The S2M can be delivered with TEDS, which is then fitted in
the transducer housing, connected and supplied with data by HBM before delivery. If the
force transducer is ordered without calibration, the characteristic values from the manu
facturing certificate are stored in the TEDS chip; if an additional DKD calibration is
ordered, the calibration results are stored in the TEDS chip.
The TEDS module is connected between PIN E (sense lead (-)) and PIN D (excitation
voltage lead (-)). HBM's zero wire configuration allows the TEDS to be read without an
additional sense lead.
If a suitable amplifier is connected (e.g. QuantumX from HBM), the amplifier electronics
will read the TEDS chip and parameterization will then be implemented automatically,
without any intervention required by the user.
The chip content can be edited and modified with suitable hardware and software. This
can be implemented, e.g. with the Quantum Assistant or even the DAQ software CATMAN
from HBM. Please follow the operating manuals of these products.
Summary of Contents for HBK S2M
Page 3: ...S2M ENGLISH DEUTSCH FRAN AIS ITALIANO Mounting Instructions...
Page 33: ...S2M ENGLISH DEUTSCH FRAN AIS ITALIANO Montageanleitung...
Page 62: ...S2M ABMESSUNGEN 30...
Page 63: ...S2M ENGLISH DEUTSCH FRAN AIS ITALIANO Notice de montage...
Page 93: ...S2M ENGLISH DEUTSCH FRAN AIS ITALIANO Istruzioni per il montaggio...
Page 123: ...31 S2M DIMENSIONI...
Page 124: ...HBK Hottinger Br el Kjaer www hbkworld com info hbkworld com A03592 01 YI0 05 7 2001 3592...