6-4
TITRATION
During the pre-titration, the user cannot change the currently selected method or access the
method parameters.
Note
: If the pre-titration lasts longer than 30 minutes the titrator switches to
Idle
mode.
Errors may have occurred in your titration system (beaker is not properly sealed,
wrong reagent, unconnected or bad electrode, etc.). Check the system and start
pre-titration again.
6.3
Drift Analysis
(
Automatic Drift Entry / KF Coulometric
only)
While in this mode, the
HI 904
conducts an automatic one-minute analysis which determines
the amount of moisture leaking into the cell from the atmosphere. Despite the titration vessel
being tightly sealed, water will still seep into the cell. The amount of water that migrates into
the cell per unit time is known as the background drift rate, or the drift rate.
The drift rate is determined by keeping track of the amount of water that has been titrated to
maintain the ‘dryness’ of the reagent over the course of a minute. The rate at which water
leaks into the cell is then calculated and reported by the
HI 904
in units of
µ
g/min.
The
HI 904
uses the drift rate determined during this state to automatically subtract the
quantity of water which leaks into the cell during a titration from titration results. This is
especially important for titration accuracy when analyzing samples with very low water
content where the amount of water which has leaked into the cell is a considerable fraction of
the total water titrated during the analysis.
When the drift becomes stable the titrator switches to
Standby
mode.
During the drift analysis, if the titrator cannot maintain cell dryness, the titrator reverts to
pre-titration.
Note
: If the drift entry mode is set as Manual, the drift analysis stage is skipped.
Summary of Contents for HI 904
Page 1: ...1 QUICK START GUIDE HI 904 KARL FISCHER COULOMETRIC TITRATOR Revision 1 0 www hannainst com...
Page 4: ...4 QUICK START GUIDE...
Page 7: ...7 QUICK START GUIDE TITRATOR CONNECTIONS Front View Rear View...
Page 17: ...1 INSTRUCTION MANUAL HI 904 KARL FISCHER COULOMETRIC TITRATOR Revision 1 00 www hannainst com...
Page 20: ...4...
Page 22: ...1 2 INTRODUCTION...
Page 24: ...2 2 SETUP...
Page 27: ...2 5 SETUP 2 3 Installation 2 3 1 Titrator Top View...
Page 28: ...2 6 SETUP 2 3 2 Titrator Rear View 2 3 3 Titrator Left side View...
Page 38: ...3 2 USER INTERFACE...
Page 46: ...3 10 USER INTERFACE...
Page 48: ...4 2 GENERAL OPTIONS...
Page 82: ...5 24 METHODS...
Page 84: ...6 2 TITRATION...
Page 94: ...6 12 TITRATION...
Page 96: ...7 2 AUXILIARY FUNCTIONS...
Page 104: ...8 2 MAINTENANCE PERIPHERALS...
Page 112: ...8 10 MAINTENANCE PERIPHERALS...
Page 124: ...9 12 METHODS OPTIMIZATION...
Page 125: ...A3 2 APPENDIX 3...
Page 126: ...A3 6 APPENDIX 3 MAN HI904 09 13...
Page 127: ...A1 1 APPENDIX 1 Appendix 1 Contents A1 TECHNICAL SPECIFICATIONS A1 3...
Page 128: ...A1 2 APPENDIX 1...
Page 132: ...A2 2 APPENDIX 2...
Page 134: ...A2 4 APPENDIX 2...
Page 135: ...A3 1 APPENDIX 3 Appendix 3 Accessories A3 TITRATOR COMPONENTS A3 3...
Page 136: ...A3 2 APPENDIX 3...
Page 140: ...A3 6 APPENDIX 3 MAN HI904 09 13...
Page 141: ...1 GENERAL APPLICATIONS BROCHURE HI 904 KARL FISCHER COULOMETRIC TITRATOR www hannainst com...
Page 142: ......
Page 144: ......
Page 146: ......
Page 150: ...2 TITRATION THEORY...
Page 166: ...18 TITRATION THEORY A B Figure 8 Potential glass electrode mV C...