5-4
METHODS
5.2
Standard Methods
The standard methods were developed for the most common types of analysis. Also, the
standard methods can be used as a model to create new user methods.
Only specific method parameters can be modified by the user (see Section
5.5,
Method
Options
section).
5.2.1
Upgrading Standard Methods
To upgrade the titrator with new standard methods, follow the steps below:
From USB Storage Device:
• Insert the USB storage device into the USB port, located on the left side of the titrator.
• Access the
General Options
screen.
• Using the
and
keys, highlight the
Restore Files from USB Storage Device
option and choose
.
• Using the
and
keys, navigate through file types to find “standard method
files”. The list with available standard methods on the storage device will be displayed.
• Press the
or
key to upgrade the titrator with the standard methods.
• Press
to return to
General Options
screen.
Note:
See section 4.8 Restore Files from USB Storage Device.
From PC:
You can upgrade the titrator with standard methods from a PC using the
HI 900
PC application
(see Section
4.9, USB Link with PC
).
5.2.2
Deleting Standard Methods
Unnecessary standard methods can be removed from titrator by following the procedure
below:
From
General Options
screen:
• Using the
and
keys, highlight the
Save Files to USB Storage Device
option
and press
;
• Using the
and
keys, navigate through the file types to find “standard method
files”. The available standard methods will be displayed.
• Press the
or
keys to remove unnecessary standard methods.
• Press
to return to the
General Options
screen.
From PC:
Unnecessary Standard Methods can be removed from the titrator using the
HI 900
PC
application (see Section
4.9, USB Link with PC
).
Summary of Contents for HI 904
Page 1: ...1 QUICK START GUIDE HI 904 KARL FISCHER COULOMETRIC TITRATOR Revision 1 0 www hannainst com...
Page 4: ...4 QUICK START GUIDE...
Page 7: ...7 QUICK START GUIDE TITRATOR CONNECTIONS Front View Rear View...
Page 17: ...1 INSTRUCTION MANUAL HI 904 KARL FISCHER COULOMETRIC TITRATOR Revision 1 00 www hannainst com...
Page 20: ...4...
Page 22: ...1 2 INTRODUCTION...
Page 24: ...2 2 SETUP...
Page 27: ...2 5 SETUP 2 3 Installation 2 3 1 Titrator Top View...
Page 28: ...2 6 SETUP 2 3 2 Titrator Rear View 2 3 3 Titrator Left side View...
Page 38: ...3 2 USER INTERFACE...
Page 46: ...3 10 USER INTERFACE...
Page 48: ...4 2 GENERAL OPTIONS...
Page 82: ...5 24 METHODS...
Page 84: ...6 2 TITRATION...
Page 94: ...6 12 TITRATION...
Page 96: ...7 2 AUXILIARY FUNCTIONS...
Page 104: ...8 2 MAINTENANCE PERIPHERALS...
Page 112: ...8 10 MAINTENANCE PERIPHERALS...
Page 124: ...9 12 METHODS OPTIMIZATION...
Page 125: ...A3 2 APPENDIX 3...
Page 126: ...A3 6 APPENDIX 3 MAN HI904 09 13...
Page 127: ...A1 1 APPENDIX 1 Appendix 1 Contents A1 TECHNICAL SPECIFICATIONS A1 3...
Page 128: ...A1 2 APPENDIX 1...
Page 132: ...A2 2 APPENDIX 2...
Page 134: ...A2 4 APPENDIX 2...
Page 135: ...A3 1 APPENDIX 3 Appendix 3 Accessories A3 TITRATOR COMPONENTS A3 3...
Page 136: ...A3 2 APPENDIX 3...
Page 140: ...A3 6 APPENDIX 3 MAN HI904 09 13...
Page 141: ...1 GENERAL APPLICATIONS BROCHURE HI 904 KARL FISCHER COULOMETRIC TITRATOR www hannainst com...
Page 142: ......
Page 144: ......
Page 146: ......
Page 150: ...2 TITRATION THEORY...
Page 166: ...18 TITRATION THEORY A B Figure 8 Potential glass electrode mV C...