9-3
METHODS OPTIMIZATION
9
OPTIMIZATION
9.1
Titration Settings
This section provides the descriptions of critical titration parameters necessary for an analyst
to modify a standard method or develop a titration method from scratch.
HI 904
methods can be modified and customized based on the requirements of the sample,
sample matrix and the Karl Fischer reagent formulation. The user changeable settings are
separated into two categories: Control Parameters, which set critical functions that determine
the course of a titration and set the way in which titrations are terminated, and Method
Options, which control lesser features not directly affecting measurements and primarily
allow advanced users to shorten titration times.
9.1.1
Control Parameters
9.1.1.1 Endpoint Potential and Polarization Current
The
HI 904
uses the polarized electrode system known as bivoltametric indication. The
titrator monitors the voltage required to maintain a pulsed polarization current (Ipol) between
the pins of a dual platinum-pin Karl Fischer electrode during the course of a titration.
During a titration, no excess iodine is present. In order to maintain the set polarization
current the
HI 904
must apply a relatively large voltage across the pins of the electrode.
At the endpoint of the titration, the amount of iodine added is equal to the amount of water
from the sample. When an excess of iodine has been generated the excess iodine is easily
reduced, and the resulting iodide is easily oxidized in electrode reactions at the cathode and
anode respectively. The ease of these reactions make maintaining the polarization current
possible at a much lower electrode potential.
In theory, a large shift in the electrode potential indicates the endpoint. In practice, a titration
endpoint is reached when the electrode potential drops below a value defined by the user
and the chosen termination criteria is met.
The choice of endpoint potential should be based, foremost, on the polarization current and,
to a lesser extent, on the composition of the Karl Fischer reagent and the sample matrix. If
the polarization current is changed, the endpoint potential must also be changed. In addition,
there are pitfalls to be avoided when choosing an endpoint potential. Selecting endpoints
which are both ‘too high’ or ‘too low’ will result in long titration times and poor reproducibility.
Endpoints which are ‘too high’ are those which result in endpoints that either precede or
coincide with equivalence point such that the concentration of excess iodine is not reliably
detected. Endpoint potentials are considered ‘too low’ when they correspond to a large
excess of iodine in the titration cell.
Summary of Contents for HI 904
Page 1: ...1 QUICK START GUIDE HI 904 KARL FISCHER COULOMETRIC TITRATOR Revision 1 0 www hannainst com...
Page 4: ...4 QUICK START GUIDE...
Page 7: ...7 QUICK START GUIDE TITRATOR CONNECTIONS Front View Rear View...
Page 17: ...1 INSTRUCTION MANUAL HI 904 KARL FISCHER COULOMETRIC TITRATOR Revision 1 00 www hannainst com...
Page 20: ...4...
Page 22: ...1 2 INTRODUCTION...
Page 24: ...2 2 SETUP...
Page 27: ...2 5 SETUP 2 3 Installation 2 3 1 Titrator Top View...
Page 28: ...2 6 SETUP 2 3 2 Titrator Rear View 2 3 3 Titrator Left side View...
Page 38: ...3 2 USER INTERFACE...
Page 46: ...3 10 USER INTERFACE...
Page 48: ...4 2 GENERAL OPTIONS...
Page 82: ...5 24 METHODS...
Page 84: ...6 2 TITRATION...
Page 94: ...6 12 TITRATION...
Page 96: ...7 2 AUXILIARY FUNCTIONS...
Page 104: ...8 2 MAINTENANCE PERIPHERALS...
Page 112: ...8 10 MAINTENANCE PERIPHERALS...
Page 124: ...9 12 METHODS OPTIMIZATION...
Page 125: ...A3 2 APPENDIX 3...
Page 126: ...A3 6 APPENDIX 3 MAN HI904 09 13...
Page 127: ...A1 1 APPENDIX 1 Appendix 1 Contents A1 TECHNICAL SPECIFICATIONS A1 3...
Page 128: ...A1 2 APPENDIX 1...
Page 132: ...A2 2 APPENDIX 2...
Page 134: ...A2 4 APPENDIX 2...
Page 135: ...A3 1 APPENDIX 3 Appendix 3 Accessories A3 TITRATOR COMPONENTS A3 3...
Page 136: ...A3 2 APPENDIX 3...
Page 140: ...A3 6 APPENDIX 3 MAN HI904 09 13...
Page 141: ...1 GENERAL APPLICATIONS BROCHURE HI 904 KARL FISCHER COULOMETRIC TITRATOR www hannainst com...
Page 142: ......
Page 144: ......
Page 146: ......
Page 150: ...2 TITRATION THEORY...
Page 166: ...18 TITRATION THEORY A B Figure 8 Potential glass electrode mV C...