
11
QUICK START GUIDE
Fill Titration Vessel with Reagent
The titration vessel must be filled with reagent up to the MIN marker (about 75 mL):
• Lower the reagent tube below the liquid level inside the reagent bottle.
• From the
Idle
screen, press
.
• Push and hold the
FILL
button located on the top of the air pump.
• Wait until the vessel is filled up to the MIN marker with solvent.
• Stop the air pump by pressing
and then confirm the approximate amount of
reagent in the vessel.
• Raise the reagent tube above the liquid in the reagent bottle.
Prepare the Reagent for Samples
Before beginning a titration, residual moisture inside the titration vessel and reagent must be
reacted:
• From the Idle screen, press
. The titrator will enter Pre-Titration mode, start the
magnetic stirrer, and begin dosing in the titration vessel.
• Once all residual moisture has been reacted (endpoint potential is reached), the
titrator will enter Drift Analysis mode (assuming Automatic Drift Entry is selected).
The titrator calculates the rate of atmospheric moisture seeping into the titration
beaker for the next minute and displays the result in the lower right corner of the
display.
• If the Drift Rate is stable and the endpoint potential is maintained, the titrator will
enter Standby mode. The titrator continues to maintain the endpoint potential and
update the background drift rate.
Note
: New (or cleaned) detector electrodes have low electrical resistance due to a lack of
platinum-iodine complexes on the electrode surface. This may cause initial mV readings to
be low and prevent proper pre-titration of the reagent. The endpoint value (100 mV by
default) should be 200-250 mV below the mV value of “wet” reagent for proper pre-titration
to occur. If necessary, adjust the endpoint and/or imposed current (in Method Options) to
facilitate proper pre-titration. The platinum-iodine complex should form after several titrations
and raise the mV readings.
Preparing and Introducing the Sample
Sample Mass Preparation
Measuring the sample size by mass using an analytical balance will give the most reproducible
results.
Summary of Contents for HI 904
Page 1: ...1 QUICK START GUIDE HI 904 KARL FISCHER COULOMETRIC TITRATOR Revision 1 0 www hannainst com...
Page 4: ...4 QUICK START GUIDE...
Page 7: ...7 QUICK START GUIDE TITRATOR CONNECTIONS Front View Rear View...
Page 17: ...1 INSTRUCTION MANUAL HI 904 KARL FISCHER COULOMETRIC TITRATOR Revision 1 00 www hannainst com...
Page 20: ...4...
Page 22: ...1 2 INTRODUCTION...
Page 24: ...2 2 SETUP...
Page 27: ...2 5 SETUP 2 3 Installation 2 3 1 Titrator Top View...
Page 28: ...2 6 SETUP 2 3 2 Titrator Rear View 2 3 3 Titrator Left side View...
Page 38: ...3 2 USER INTERFACE...
Page 46: ...3 10 USER INTERFACE...
Page 48: ...4 2 GENERAL OPTIONS...
Page 82: ...5 24 METHODS...
Page 84: ...6 2 TITRATION...
Page 94: ...6 12 TITRATION...
Page 96: ...7 2 AUXILIARY FUNCTIONS...
Page 104: ...8 2 MAINTENANCE PERIPHERALS...
Page 112: ...8 10 MAINTENANCE PERIPHERALS...
Page 124: ...9 12 METHODS OPTIMIZATION...
Page 125: ...A3 2 APPENDIX 3...
Page 126: ...A3 6 APPENDIX 3 MAN HI904 09 13...
Page 127: ...A1 1 APPENDIX 1 Appendix 1 Contents A1 TECHNICAL SPECIFICATIONS A1 3...
Page 128: ...A1 2 APPENDIX 1...
Page 132: ...A2 2 APPENDIX 2...
Page 134: ...A2 4 APPENDIX 2...
Page 135: ...A3 1 APPENDIX 3 Appendix 3 Accessories A3 TITRATOR COMPONENTS A3 3...
Page 136: ...A3 2 APPENDIX 3...
Page 140: ...A3 6 APPENDIX 3 MAN HI904 09 13...
Page 141: ...1 GENERAL APPLICATIONS BROCHURE HI 904 KARL FISCHER COULOMETRIC TITRATOR www hannainst com...
Page 142: ......
Page 144: ......
Page 146: ......
Page 150: ...2 TITRATION THEORY...
Page 166: ...18 TITRATION THEORY A B Figure 8 Potential glass electrode mV C...