6-3
TITRATION
6
TITRATION
6.1
Idle
The titrator first enters Idle mode when it is switched on. All of the
HI 904
’s software features
and adjustable parameters can be accessed from the Idle state. This includes all of the
user-adjustable method parameters, solvent handling system, file transfers, calibration checks,
software upgrades, options for interface with PC and accessories as well as burette options.
To access the titration menu (
Process
screen) press
.
The titration is performed with the selected method.
Be sure that the selected method is customized in accordance with the specifics of the application.
Before performing a titration, make sure that the following conditions are met:
• All of the attached systems (e.g.: reagent system) are properly assembled.
• The right amount of reagent is present in the titration vessel (between the min and
max marks) for best reproducibility.
The following intermediary stages are performed automatically before starting the analysis:
•
Reagent pre-titration
•
Drift analysis
(
Automatic Drift Entry / KF Coulometric
only)
When the drift analysis process is finished, the titrator enters
Standby
mode. At this point,
a titration can be initiated.
6.2
Pre-titration
In pre-titration, the residual water on the interior surface of the titration vessel, the water
contained in the entrapped air and the small amount of water from the reagent, is elimi-
nated. The
HI 904
reacts residual water with iodine that is generated electrolytically inside
the titration vessel until the specified end point potential is reached; this setting is associated
with the active method. After the electrode potential has stabilized, the titrator moves into
the Drift Rate Determination Stage.
When the pre-titration is started, the stirrer is automatically turned on (when
Internal
or
External
stirrer is selected).
Summary of Contents for HI 904
Page 1: ...1 QUICK START GUIDE HI 904 KARL FISCHER COULOMETRIC TITRATOR Revision 1 0 www hannainst com...
Page 4: ...4 QUICK START GUIDE...
Page 7: ...7 QUICK START GUIDE TITRATOR CONNECTIONS Front View Rear View...
Page 17: ...1 INSTRUCTION MANUAL HI 904 KARL FISCHER COULOMETRIC TITRATOR Revision 1 00 www hannainst com...
Page 20: ...4...
Page 22: ...1 2 INTRODUCTION...
Page 24: ...2 2 SETUP...
Page 27: ...2 5 SETUP 2 3 Installation 2 3 1 Titrator Top View...
Page 28: ...2 6 SETUP 2 3 2 Titrator Rear View 2 3 3 Titrator Left side View...
Page 38: ...3 2 USER INTERFACE...
Page 46: ...3 10 USER INTERFACE...
Page 48: ...4 2 GENERAL OPTIONS...
Page 82: ...5 24 METHODS...
Page 84: ...6 2 TITRATION...
Page 94: ...6 12 TITRATION...
Page 96: ...7 2 AUXILIARY FUNCTIONS...
Page 104: ...8 2 MAINTENANCE PERIPHERALS...
Page 112: ...8 10 MAINTENANCE PERIPHERALS...
Page 124: ...9 12 METHODS OPTIMIZATION...
Page 125: ...A3 2 APPENDIX 3...
Page 126: ...A3 6 APPENDIX 3 MAN HI904 09 13...
Page 127: ...A1 1 APPENDIX 1 Appendix 1 Contents A1 TECHNICAL SPECIFICATIONS A1 3...
Page 128: ...A1 2 APPENDIX 1...
Page 132: ...A2 2 APPENDIX 2...
Page 134: ...A2 4 APPENDIX 2...
Page 135: ...A3 1 APPENDIX 3 Appendix 3 Accessories A3 TITRATOR COMPONENTS A3 3...
Page 136: ...A3 2 APPENDIX 3...
Page 140: ...A3 6 APPENDIX 3 MAN HI904 09 13...
Page 141: ...1 GENERAL APPLICATIONS BROCHURE HI 904 KARL FISCHER COULOMETRIC TITRATOR www hannainst com...
Page 142: ......
Page 144: ......
Page 146: ......
Page 150: ...2 TITRATION THEORY...
Page 166: ...18 TITRATION THEORY A B Figure 8 Potential glass electrode mV C...