TITRATION THEORY
The amount of current that passes through the generator can easily and accurately be measured
by the electronics of the titrator. Therefore, coulometric Karl Fischer titrations are considered
absolute - standardization is not necessary. Water standards can be titrated as a system check
to ensure proper system functioning.
2.2.1.3.1
Generator Electrodes with Diaphragm
The first coulometric Karl Fischer titrators used a diaphragm cell. In this design, the anode
and cathode of the generator are separated by a diaphragm made typically of porous frit
glass. The diaphragm serves to prevent the iodine generated at the anode from being reduced
at the cathode, which would cause false high water determinations. The anode compartment
contains the Karl Fischer reaction components (sulfur dioxide, methanol, base) and iodide
salts for the generation of molecular iodine. The cathode compartment contains a source of
hydrogen ions, typically ammonium salts.
Diaphragm titrations have some disadvantages. The first disadvantage is the higher drift
rates that occur due to moisture collecting inside the catholyte. Since the Karl Fischer reaction
only occurs in the anode compartment, moisture inside the catholyte cannot be eliminated
by pre-titration. Instead of being pre-titrated, the moisture inside that catholyte will slowly
diffuse across the diaphragm during drift analysis and sample analysis, and will add to the
apparent drift rate. The second disadvantage is the risk of diaphragm blockage or
contamination. Substances in the sample matrix may clog the diaphragm, or salts could
precipitate inside the diaphragm. A clogged diaphragm will prevent ion migration which, in
severe cases, will block the electrolytic reaction of the generator. The third disadvantage is
difficulty in cleaning. The diaphragm does not absorb or drain fluid quickly, making cleaning
very time-consuming. The cathode compartment itself is also not very accessible for cleaning.
2.2.1.3.2
Generator Electrodes without Diaphragm
To overcome the drawbacks of diaphragm titrations, diaphragm-less titration systems were
made through modification of the generator’s design and modification of the reagent. The
cathode’s surface is much smaller compared to the anode, allowing the generated iodine to
react before possibly reaching the cathode. The reagent is also modified to prevent oxidizable
sulfur compounds from forming.
Diaphragm-less titration offers very low drift rates and easy cell maintenance, but there are
several drawbacks. First, side reactions are prone to occur particularly at slower titration
rates. Therefore, samples with very low water contents may suffer from false high
concentrations. Second, compounds that are easily reduced will react at the cathode and
produce water, causing false high concentrations. These compounds include nitro compounds,
unsaturated hydrocarbons, and certain metals.
Summary of Contents for HI 904
Page 1: ...1 QUICK START GUIDE HI 904 KARL FISCHER COULOMETRIC TITRATOR Revision 1 0 www hannainst com...
Page 4: ...4 QUICK START GUIDE...
Page 7: ...7 QUICK START GUIDE TITRATOR CONNECTIONS Front View Rear View...
Page 17: ...1 INSTRUCTION MANUAL HI 904 KARL FISCHER COULOMETRIC TITRATOR Revision 1 00 www hannainst com...
Page 20: ...4...
Page 22: ...1 2 INTRODUCTION...
Page 24: ...2 2 SETUP...
Page 27: ...2 5 SETUP 2 3 Installation 2 3 1 Titrator Top View...
Page 28: ...2 6 SETUP 2 3 2 Titrator Rear View 2 3 3 Titrator Left side View...
Page 38: ...3 2 USER INTERFACE...
Page 46: ...3 10 USER INTERFACE...
Page 48: ...4 2 GENERAL OPTIONS...
Page 82: ...5 24 METHODS...
Page 84: ...6 2 TITRATION...
Page 94: ...6 12 TITRATION...
Page 96: ...7 2 AUXILIARY FUNCTIONS...
Page 104: ...8 2 MAINTENANCE PERIPHERALS...
Page 112: ...8 10 MAINTENANCE PERIPHERALS...
Page 124: ...9 12 METHODS OPTIMIZATION...
Page 125: ...A3 2 APPENDIX 3...
Page 126: ...A3 6 APPENDIX 3 MAN HI904 09 13...
Page 127: ...A1 1 APPENDIX 1 Appendix 1 Contents A1 TECHNICAL SPECIFICATIONS A1 3...
Page 128: ...A1 2 APPENDIX 1...
Page 132: ...A2 2 APPENDIX 2...
Page 134: ...A2 4 APPENDIX 2...
Page 135: ...A3 1 APPENDIX 3 Appendix 3 Accessories A3 TITRATOR COMPONENTS A3 3...
Page 136: ...A3 2 APPENDIX 3...
Page 140: ...A3 6 APPENDIX 3 MAN HI904 09 13...
Page 141: ...1 GENERAL APPLICATIONS BROCHURE HI 904 KARL FISCHER COULOMETRIC TITRATOR www hannainst com...
Page 142: ......
Page 144: ......
Page 146: ......
Page 150: ...2 TITRATION THEORY...
Page 166: ...18 TITRATION THEORY A B Figure 8 Potential glass electrode mV C...