20
TITRATION THEORY
3.2 Automatic Titration
Automatic titrators are high-precision analytical instruments that deliver the titrant, monitor
the physical change associated with the titration reaction, automatically stops at the endpoint
and calculates the concentration of the analyte. Automatic titrators are best for repetitive
titrations and high-accuracy analyses.
An automatic titrator must have an accurate liquid dispensing system. In high accuracy
systems like the
HI 900
-series titrators, the liquid dispensing system consists of a stepper-motor
driven piston syringe burette capable of accurately and precisely dispensing very small volumes
of titrant, a valve system to switch between titrant intake and outlet and an anti-diffusion
dispensing tip. These three main subsystem components must be as accurate as possible,
with very low gear backlash in the burette pump, minimal piston seal flexing, precision
ground inner diameter of the glass syringe, a low dead volume valve, minimal evaporation/
permeation, and chemically resistant tubing.
Apparatus required for automatic titration include:
• An automatic titrator, equipped with a burette;
• A beaker;
• An electronic stirring system, either a propeller stirrer or a magnetic stir bar and stir
plate;
• Volumetric pipettes for the precise addition of samples;
• Standard titrant solutions of known concentration;
• An electrode system that can be used to determine the endpoint of the titration.
A typical automatic titration consists of the following steps:
1. Set up the automatic titrator according to the manufacturer’s instructions;
2. A volumetric pipette is typically used to add a known volume of sample to the beaker;
3. Submerge the propeller stirrer or add the stir bar to the beaker, and turn on;
4. Start the titration, the titrator will automatically stop at the endpoint and determine
the concentration of the analyte.
Summary of Contents for HI 904
Page 1: ...1 QUICK START GUIDE HI 904 KARL FISCHER COULOMETRIC TITRATOR Revision 1 0 www hannainst com...
Page 4: ...4 QUICK START GUIDE...
Page 7: ...7 QUICK START GUIDE TITRATOR CONNECTIONS Front View Rear View...
Page 17: ...1 INSTRUCTION MANUAL HI 904 KARL FISCHER COULOMETRIC TITRATOR Revision 1 00 www hannainst com...
Page 20: ...4...
Page 22: ...1 2 INTRODUCTION...
Page 24: ...2 2 SETUP...
Page 27: ...2 5 SETUP 2 3 Installation 2 3 1 Titrator Top View...
Page 28: ...2 6 SETUP 2 3 2 Titrator Rear View 2 3 3 Titrator Left side View...
Page 38: ...3 2 USER INTERFACE...
Page 46: ...3 10 USER INTERFACE...
Page 48: ...4 2 GENERAL OPTIONS...
Page 82: ...5 24 METHODS...
Page 84: ...6 2 TITRATION...
Page 94: ...6 12 TITRATION...
Page 96: ...7 2 AUXILIARY FUNCTIONS...
Page 104: ...8 2 MAINTENANCE PERIPHERALS...
Page 112: ...8 10 MAINTENANCE PERIPHERALS...
Page 124: ...9 12 METHODS OPTIMIZATION...
Page 125: ...A3 2 APPENDIX 3...
Page 126: ...A3 6 APPENDIX 3 MAN HI904 09 13...
Page 127: ...A1 1 APPENDIX 1 Appendix 1 Contents A1 TECHNICAL SPECIFICATIONS A1 3...
Page 128: ...A1 2 APPENDIX 1...
Page 132: ...A2 2 APPENDIX 2...
Page 134: ...A2 4 APPENDIX 2...
Page 135: ...A3 1 APPENDIX 3 Appendix 3 Accessories A3 TITRATOR COMPONENTS A3 3...
Page 136: ...A3 2 APPENDIX 3...
Page 140: ...A3 6 APPENDIX 3 MAN HI904 09 13...
Page 141: ...1 GENERAL APPLICATIONS BROCHURE HI 904 KARL FISCHER COULOMETRIC TITRATOR www hannainst com...
Page 142: ......
Page 144: ......
Page 146: ......
Page 150: ...2 TITRATION THEORY...
Page 166: ...18 TITRATION THEORY A B Figure 8 Potential glass electrode mV C...