21
TITRATION THEORY
4
TITRATION RESULTS
4.1 Accuracy
The factors most critical to achieving accurate results with the
HI 900
titration systems are
the concentration of the sample, size of the sample and having an optimized set of method
parameters.
4.2 Repeatability
Repeatability, or the agreement between replicate determinations, is expressed quantitatively
as the relative standard deviation (RSD).
4.3 Sources of Error
One of the advantages of volumetric analysis is excellent accuracy and precision. The sources
of error can be grouped into sampling, titrant and standards, chemical reactions, endpoint
determination and calculations.
4.3.1 Sampling Errors
• Selection of a non-homogeneous or non-representative sample;
• Sample changed or was contaminated during collection, storage or transfers;
• Poor technique when transferring sample to beaker or flask;
• Errors in the balance, calibrate and check balance regularly.
4.3.2 Errors with Titrant and Standard
4.3.2.1
Preparation Errors
Incorrect preparation due to:
• Poor technique in weighing the salt or when transferring to volumetric glassware;
• Low-purity of salts or water used to make titrant and standard;
• Dirty or wet glassware;
• Improper storage of titrant or standard which allows water gain, evaporation or
deterioration;
• Failure to standardize frequently to adjust for change in titrant;
• Failure to flush titrator tubing with a volume of titrant before standardizing;
• Volume errors from pipettes and volumetric flasks, grade A glassware is required;
• Balance errors when weighing out salts, calibrate and check balance regularly.
Summary of Contents for HI 904
Page 1: ...1 QUICK START GUIDE HI 904 KARL FISCHER COULOMETRIC TITRATOR Revision 1 0 www hannainst com...
Page 4: ...4 QUICK START GUIDE...
Page 7: ...7 QUICK START GUIDE TITRATOR CONNECTIONS Front View Rear View...
Page 17: ...1 INSTRUCTION MANUAL HI 904 KARL FISCHER COULOMETRIC TITRATOR Revision 1 00 www hannainst com...
Page 20: ...4...
Page 22: ...1 2 INTRODUCTION...
Page 24: ...2 2 SETUP...
Page 27: ...2 5 SETUP 2 3 Installation 2 3 1 Titrator Top View...
Page 28: ...2 6 SETUP 2 3 2 Titrator Rear View 2 3 3 Titrator Left side View...
Page 38: ...3 2 USER INTERFACE...
Page 46: ...3 10 USER INTERFACE...
Page 48: ...4 2 GENERAL OPTIONS...
Page 82: ...5 24 METHODS...
Page 84: ...6 2 TITRATION...
Page 94: ...6 12 TITRATION...
Page 96: ...7 2 AUXILIARY FUNCTIONS...
Page 104: ...8 2 MAINTENANCE PERIPHERALS...
Page 112: ...8 10 MAINTENANCE PERIPHERALS...
Page 124: ...9 12 METHODS OPTIMIZATION...
Page 125: ...A3 2 APPENDIX 3...
Page 126: ...A3 6 APPENDIX 3 MAN HI904 09 13...
Page 127: ...A1 1 APPENDIX 1 Appendix 1 Contents A1 TECHNICAL SPECIFICATIONS A1 3...
Page 128: ...A1 2 APPENDIX 1...
Page 132: ...A2 2 APPENDIX 2...
Page 134: ...A2 4 APPENDIX 2...
Page 135: ...A3 1 APPENDIX 3 Appendix 3 Accessories A3 TITRATOR COMPONENTS A3 3...
Page 136: ...A3 2 APPENDIX 3...
Page 140: ...A3 6 APPENDIX 3 MAN HI904 09 13...
Page 141: ...1 GENERAL APPLICATIONS BROCHURE HI 904 KARL FISCHER COULOMETRIC TITRATOR www hannainst com...
Page 142: ......
Page 144: ......
Page 146: ......
Page 150: ...2 TITRATION THEORY...
Page 166: ...18 TITRATION THEORY A B Figure 8 Potential glass electrode mV C...