4
ADVANCED OPERATION EXAMPLE
APS-1102A
4-5
4.1.3 Setting the output voltage range
Select 100 V or 200 V as the output voltage range.
As is shown in
Table4-2
, the range that can be set for the output voltage differs according to the
selected output voltage range.
Table4-2. Setting Range Options for Various Output Voltage Ranges
To select an output voltage range, select
in the SET menu.
The output voltage range cannot be changed when output is on.
Operation steps
1.
Press the
or
key to move the cursor to the
icon.
ENTER
CANCEL
Cursor
2.
Press the
EN TER
key to display the selection box.
ENTER
CANCEL
Selection box
Setting
Setting range
100 V range
200 V range
Unit
DC voltage
-220.0 to +220.0
-440.0 to +440.0
V
AC voltage
SIN/SQU
0.0 to 155.0
0.0 to 310.0
Vrms
ARB1 to ARB16
0.0 to 440.0
0.0 to 880.0
Vp-p
Peak current limiter (positive)
+10.0 to +42.0
+5.0 to +21.0
A
Peak current limiter (negative)
-42.0 to -10.0
-21.0 to
-5.0
A
RMS current limiter
1.0 to
10.5
1.0 to
5.3
A
Positive voltage setting limit
+0.1 to +220.0
+0.1 to +440.0
V
Negative voltage setting limit
-220.0 to
-0.1
-440.0 to
-0.1
V
External input gain
0.0 to 220.0
0.0 to 440.0
times