4
ADVANCED OPERATION EXAMPLE
APS-1102A
4-47
c) Operations after starting sequence
When a sequence is started, the status icon (
) is displayed in the status window to indicate that
the sequence is being executed. A message box is also displayed to indicate the step number being
executed.
After returning to the SET screen, if the sequence is executed, the current step settings will not be
shown in the SET screen. When the sequence is completed, the final step settings will be shown.
If the
final step voltage settings are outside of the setting range limit, both DC voltage and AC voltage will
be set to 0 V. If the final step frequency settings are outside of the setting range limit, frequency will be
set to the limit value of the setting range limit. In either case, the output will be turned off.
Status icon indicates that sequence
is being executed
Step number being executed
Figure 4-14. LCD Screen During Sequence Execution