9
SPECIFICATIONS
APS-1102A
9-9
Average value:
Average value of AC+DC (measures DC component)
Full scale:
15.00 A
Display resolution:
0.01 A
Measurement accuracy:
(|0.5% of rdg| + 0.08 A)
(at 23
5°C, output current is 5% to 100% of the maximum
current)
Peak value:
Minimum current and maximum current are displayed
separately
Full scale:
45.0 Apk
Display resolution:
0.1 Apk
Measurement accuracy:
(|2% of rdg| + 0.4 Apk)
(reference value, at 23
5°C, sine wave of which frequency is 45
Hz to 65 Hz)
Peak value hold:
Maximum value of
max current
and
min current |
Full scale:
45.0 Apk
Display resolution:
0.1 Apk
c) Output power measurement
Effective power
Full scale:
1200 W
Display resolution:
1 W
Measurement accuracy:
at 45 Hz to 65 Hz
(2% of rdg + 1 W)
at DC
(3% of rdg + 12 W)
(at 23
5°C, load with the power factor 0.5 to lower than 1, 50 V
or higher output voltage, output current is 10% to 100% of the
maximum current)
Apparent power:
Calculated as output voltage RMS value × output current RMS
value
Full scale:
1400 VA
Display resolution:
1 VA
Measurement accuracy:
at 45 Hz to 65 Hz
(2% of rdg + 1 VA)
at DC
(3% of rdg + 12 VA)
(at 23
5°C, 50 V or higher output voltage, output current is
10% to 100% of the maximum current)