3-24
3.
Press the
or
key to move the cursor to the waveform to be output.
ENTER
CANCEL
4.
Press the
EN TER
key to set the selected waveform.
ENTER
CANCEL
3.4.5 Setting output voltage
For AC-INT mode, the output voltage settings listed below can be selected.
Table3-10. Output Voltage Settings
Output Mode
Setting
Setting Range
Resolution
AC-INT
(AC mode)
AC
voltage
100 V range
SIN/SQU
0.0 to 155.0 Vrms
0.1 Vrms
ARB1 to ARB 16
0.0 to 440.0 Vp-p
0.1 Vp-p
200 V range
SIN/SQU
0.0 to 310.0 Vrms
0.1 Vrms
ARB1 to ARB 16
0.0 to 880.0 Vp-p
0.1 Vp-p
S
ee “3.4.7 Using limiter functions”
, when using a setting range limit for output voltage.
See
“5.3.3 Setting output voltage”
, for output voltage setting ranges for other output modes.