the PEEP valve.
Adjust the PEEP cap to set the desired PEEP level as displayed on the
manometer.
8
Remove the tst lung from the T-piece. Fit the desired neonatal resuscitation mask
or endotracheal tube to the T-piece. The Neopuff resuscitator is now ready for
use.
9
Place the mask over the baby’s mouth and nose, or insert the endotracheal tube
into de baby’s airway.
10 To
resuscitate:
Place your thumb over the PEEP cap aperture on the T-piece to produce an
inspiration.
Remove your thumb from the PEEP cap aperture on the T-piece to allow an
expiration.
Cleaning.
Ensure all oxygen and air supplies are turned off and disconnected from the Neopuff
before performing cleaning procedures.
Clean the Neopuff and accessories either weekly or between babies.
Do not use solvents, alcohol-based, or abrasive cleaning solutions.
Ensure all parts and accessories are checked before returning the device to service.
For ethylene oxide gases: Some carrier gases can cause stress craking and are not
Neopuff
Test lung
Mask
Supply tube
T-piece
Steam autoclavable
136 ºC / 220 kPa / 4 minutes
✔
✔
✔
Steam autoclavable
120 ºC / 96 kPa / 15 minutes
✔
✔
✔
Ethylene Oxide Gas
✔
✔
✔
✔
Chemical liquid immersion
✔
✔
Wipe with soft damp cloth
✔
Summary of Contents for NESTOMAT 6050
Page 31: ...GUIDO RAYOS X S A NESTOMAT 6050 31 MAN 012 Sept 00 Ed 1 Rev 3 1 2 2 3 4 5 6 7 8 9 10...
Page 37: ...GUIDO RAYOS X S A NESTOMAT 6050 37 MAN 012 Sept 00 Ed 1 Rev 3...
Page 38: ...GUIDO RAYOS X S A NESTOMAT 6050 38 MAN 012 Sept 00 Ed 1 Rev 3...
Page 47: ...GUIDO RAYOS X S A NESTOMAT 6050 47 MAN 012 Sept 00 Ed 1 Rev 3...