NOTE: ST
MEANS PRESENT TEMPERATURE
. T
EMPERATURE IN BRACKETS
MEANS PRESET TEMPERATURE
.
Then press
START
to start the heating. The heat lamp will indicate the
operation of the heating device. By reaching the selected patient
temperature, the heating device will modulate to keep the temperature
stable.
The LCD will inform the patient temperature and the selected one. By
displaying
UNIT ACTIVE
will inform the heating device is on
operation.
The set patient temperature can be modified at any time by pressing
START/STOP
and after
ENTER
key; you will find displayed following
message :
AIR=__._ ºC
SELECT MODE
Then set the new temperature by the keypad as previously explained and
confirm by pressing
ENTER
key. The LCD will display the messages
described above. Then press
START
key.
ATTENTION: IF A MOMENTARY PAUSE WITHOUT
CHANGING THE SET PATIENT TEMPERATURE IS
REQUIRED, PRESS STOP KEY. PRESS START KEY TO
CONTINUE OPERATION. THE RADIANT WARMER WILL
OPERATE ON THE SAME OPERATION MODE AND
PARAMETERS ALREADY SET.
II.4 TEST TIMER.
This timer has a visual and audible alarms after 1, 5 and 10 minutes (1 + 4 + 5) since it
is switched on.
To operate the timer:
•
Press I/O key to activate TEST TIMER device.
•
Press START/STOP key to iniciate first cycle (1 minute).
•
When the display shows 1 minute, the device will stop and alrmas (visual and
audible) will be activated.
•
Press START/STOP key to silence alarms and begin next cycle (4 minutes).
•
When the display shows 5 minutes, the device will stop and alrmas (visual and
audible) will be activated.
•
Press START/STOP key to silence alarms and begin last cycle (5 minutes).
•
When the display shows 10 minutes, the device will stop and alrmas (visual and
audible) will be activated. Press START/STOP key to silence alarms; the device
will be disconected automatically.
Summary of Contents for NESTOMAT 6050
Page 31: ...GUIDO RAYOS X S A NESTOMAT 6050 31 MAN 012 Sept 00 Ed 1 Rev 3 1 2 2 3 4 5 6 7 8 9 10...
Page 37: ...GUIDO RAYOS X S A NESTOMAT 6050 37 MAN 012 Sept 00 Ed 1 Rev 3...
Page 38: ...GUIDO RAYOS X S A NESTOMAT 6050 38 MAN 012 Sept 00 Ed 1 Rev 3...
Page 47: ...GUIDO RAYOS X S A NESTOMAT 6050 47 MAN 012 Sept 00 Ed 1 Rev 3...