Ed. 1 / Rev. 3
AT=__._ ºC (__._)
UNIT ACTIVE
The selected heating range can be modified at any time by pressing
START/STOP
and after
ENTER
key; you will find displayed following
message :
AIR=__._ ºC
SELECT MODE
Then set the new temperature by the keypad as previously explained and
confirm by pressing
ENTER
key. The LCD will display the messages
described above. Then press
START
key.
B) SERVO CONTROL MODE:
ATTENTION: BEFORE SELECTING SERVO CONTROL MODE
PROCEED TO PREHEAT THE RADIANT WARMER AS
EXPLAINED PREVIOUSLY IN THE II.2. TO PREHEAT
CHAPTER.
Before anything display on the screen following message :
AIR=__._ ºC
SELECT MODE
Press SERVO key, the LCD will display ,
SKIN=__._ ºC
PSET SKIN=__._
Set the patient temperature between 20 and 40 ºC by means of the
keypad, confirm by pressing ENTER key.
NOTE: W
HEN SETTING TEMPERATURE ENTER ALWAYS
3
DIGITS
,
EVEN IN
CASE OF NO DECIMALS
. A
S EXAMPLE PRESS
3,
THEN
7,
THEN
0
TO SET
37
DEGRESS
.
The LCD will display ,
ST=__._ ºC (__._)
PRESS START