GUIDO RAYOS X, S.A.
NESTOMAT 6050
19
MAN-012
Sept.00
Ed. 1 / Rev. 3
C) POWER FAILURE
Is activated when failure of the power from the mains. The alarm circuit
includes a rechargeable battery and operation conditions of it must be
checked periodically.
To switch off the alarm operation, if desired, refer to chapter
II.6.
"
SENSORS ACTIVATION AND DEACTIVATION
".
II.7. SENSORS ACTIVATION AND DEACTIVATION
The applications program of the NESTOMAT 6050 Radiant warmer allows to the user,
at any moment, to deactivate or activate the sensors of the monitored parameters, if
required for calibration procedures or non essential parameters cancellation (Oxygen
and/or Relative Humidity).
If the Radiant warmer is switched off, proceed as follows:
Press
I/O
switch, located at the side of the trolley.
Press
ON/OFF
key at the front panel. The LCD will display ,
GUIDO RAYOS X
VER X.X
then type sequentially:
•
SILENCE
key
•
7388
•
SILENCE
key.
•
ENTER
then the LCD will display ,
MAINT: 0 SW, 1 LNG
2 SEN, 3 DIS, 4 RNG
Press
2
key and the LCD will display all coded available sensors and their status
as
follows,
1RH 2AT 3ST 4HT
5FA 6OX 7A2 8AI
(at the right side of each coded sensor "
+
" or "
-
" will display, meaning activated or
deactivated respectively)
(RH=Relative Humidity - Non operative, AT=Air Temperature, ST=Skin
Temperature,HT=Heating Device Temperature, FA=Fan - Non operative, OX=Oxygen
- Optional, A2 and A1 are auxiliary options for future implements)
To activate (+) or deactivate (-) press the corresponding figure key of the sensor.
Proceed with any sensor you want to activate or deactivate, then press ENTER to
Summary of Contents for NESTOMAT 6050
Page 31: ...GUIDO RAYOS X S A NESTOMAT 6050 31 MAN 012 Sept 00 Ed 1 Rev 3 1 2 2 3 4 5 6 7 8 9 10...
Page 37: ...GUIDO RAYOS X S A NESTOMAT 6050 37 MAN 012 Sept 00 Ed 1 Rev 3...
Page 38: ...GUIDO RAYOS X S A NESTOMAT 6050 38 MAN 012 Sept 00 Ed 1 Rev 3...
Page 47: ...GUIDO RAYOS X S A NESTOMAT 6050 47 MAN 012 Sept 00 Ed 1 Rev 3...