GUIDO RAYOS X, S.A.
NESTOMAT 6050
15
MAN-012
Sept.00
Ed. 1 / Rev. 3
II.5 OXYGEN MONITORING (OPTIONAL).
NESTOMAT 6050 Radiant Warmer monitors the oxygen percentage (optional) when
used with a tent hood (optional).
To set the alarm level for above mentioned parameter, proceed as follows (do the
calibration with the sensor out of the tent hood, as it has to be done at a concentration of
oxygen of 21%, which is the oxygen concentration of air) :
N
OTE
: W
HEN SELECTING YOU MUST START FROM
“
SELECT MODE
”
SCREEN
.
AIR=__._ ºC
SELECT MODE
Press the
O2/HUM
key, the LCD will display ,
AIR=__._ ºC
1=OXYG, 2=HUMID
I
MPORTATNT
: O
PTION
2=H
UMID IS NOT OPERATIVE FOR
R
ADIANT
W
ARMERS
.
To set the Oxygen range press
1
, the LCD will display
AIR=__._ ºC
OX: 1=PRG, 2=CAL
Then, press
1
again ,
AT=__._ ºC (__._)
PRESET OX %=21.0
By means of the keypad set the desired value. Press
ENTER
key to
confirm.
Once again you will find following screen :
AIR=__._ ºC
1=OXYG, 2=HUMID
Finally press
ENTER
to confirm values.
To calibrate the Oxygen Sensor
, press the
O2/HUM
. key, the LCD will
display ;
AIR=__._ ºC
Summary of Contents for NESTOMAT 6050
Page 31: ...GUIDO RAYOS X S A NESTOMAT 6050 31 MAN 012 Sept 00 Ed 1 Rev 3 1 2 2 3 4 5 6 7 8 9 10...
Page 37: ...GUIDO RAYOS X S A NESTOMAT 6050 37 MAN 012 Sept 00 Ed 1 Rev 3...
Page 38: ...GUIDO RAYOS X S A NESTOMAT 6050 38 MAN 012 Sept 00 Ed 1 Rev 3...
Page 47: ...GUIDO RAYOS X S A NESTOMAT 6050 47 MAN 012 Sept 00 Ed 1 Rev 3...