Press Alarm Selec. to step through or set each of the four alarm limits.
Pressing a fifth time sets the monitor at none selected. If neither the
Up/Down arrows nor the Alarm Selec. key are pressed for aproximately
20 seconds, the system returns to none selected.
8) Up/Down Arrows
When an alarm limit is not selected, these keys increase/decrease either
the alarm volume (when the alarm is not silenced) or the pulse "beep"
volume (when the alarm is silenced). When an alarm limit is selected,
these keys control scrolling up and down through the alarms limits
setting.
9) Alarm Silence
Pressing this key turns the audible alarm off either for two minutes or
until the key is pressed again. Pressing and holding it for 3 seconds turns
the alarm off until the key is pressed again or the I/O is turned off and on.
10) Alarm LED
Indicates a patient or system alarm. The LED flashes or remains steady
depending on the alarm condition (See Alarms and Indicators Chapter).
Alarms and Indicators
1) Patient:
- Pulse beat detection:
A short "beep" sounds each time a pulse beat is
detected. The volume is adjustable (including off) independently from the
alarm volume.
- Matched or Exceeded Alarm Limit:
A two-tone alarm sounds (when the
alarm is not silenced), the red Alarm LED flashes, and the triangular
indicator for the violated alarm limit flashes.
- Low Pulse Amplitud:
LOW PULSE is displayed.
- Drop in SaO2:
A lower-tone "beep" sounds.
2) System:
-
Probe Off patient or not connected to unit:
A double "beep" sounds
every second (when the alarm is not silenced), CHECK SENSOR is
displayed and the red alarm LED lights continuously.
- Searching too long for Pulse:
A double "beep" sounds every second
(when the alarm is not silenced), PULSE SEARCH is displayed and the
red alarm LED lights continuously.
Starting Pulseoximetry
WARNING: O
PERATION OF
P
ULSEOXIMETRY MAY BE AFFECTED IN THE
Summary of Contents for NESTOMAT 6050
Page 31: ...GUIDO RAYOS X S A NESTOMAT 6050 31 MAN 012 Sept 00 Ed 1 Rev 3 1 2 2 3 4 5 6 7 8 9 10...
Page 37: ...GUIDO RAYOS X S A NESTOMAT 6050 37 MAN 012 Sept 00 Ed 1 Rev 3...
Page 38: ...GUIDO RAYOS X S A NESTOMAT 6050 38 MAN 012 Sept 00 Ed 1 Rev 3...
Page 47: ...GUIDO RAYOS X S A NESTOMAT 6050 47 MAN 012 Sept 00 Ed 1 Rev 3...