GUIDO RAYOS X, S.A.
NESTOMAT 6050
18
MAN-012
Sept.00
Ed. 1 / Rev. 3
Informs when a failure of the skin probe or if it gets disconnected or out
of range. The Alarm will activate the corresponding LED and buzzer.
The LCD will display ,
* ALARM *
SKIN SENSOR
being possible to silence. The Radiant warmer is now out of service. The
Radiant warmer can be used on Air Mode if no any circunstances or
alarms avoid it, in case it is absolutely needed switch to Air Operation
Mode and call to the authorized service engineer.
2. LOW TEMPERATURE
Is activated when temperature reaches aproximately 1 ºC below the preset
temperature (See paragrap II.10 to change preset temperature alarm
range). "LOW" alarm LED will lit and the LCD will display,
* ALARM *
ST - RH OX
to silence acoustic alarm press SILENCE key. If SILENCE was pressed
and the alarm conditions remain during a period of ten minutes, the alarm
will be activated again.
IMPORTANT: C
HECK THE CONDITIONS OF THE PATIENT
. C
HECK THIERE IS
NOT AIR FLOW IN THE WARD
. C
HECK THE
HEAT L
AMP IS ON
(
THE
MICROPROCESSOR CONTROLS THE HEATING DEVICES
);
IF CONTRARY
,
UNPLUG
THE UNIT FROM THE POWER SOURCE
,
REMOVE THE PATIENT FROM THE
RADIANT WARMER
AND CALL TO THE AUTHORISED SERVICE ENGINEER
.
3. HIGH TEMPERATURE
Is activated when the temperature reaches aproximately 1 ºC higher than
the preset temperature (See paragrap II.10 to change preset temperature
alarm range). "HIGH" alarm LED will lit and the LCD will display ,
* ALARM *
ST + RH OX
to silence acoustic alarm press SILENCE key. HEAT lamp will switch
off. If SILENCE was pressed and the alarm conditions remain during a
period of ten minutes, the alarm will be activated again.
Check the conditions of the patient. Check if heat sources are close to the
Radiant warmer location, if positive keep them away.
If the alarm is still effective, unplug the unit from the power source,
remove the patient from the radiant warmer and call to the authorised
service engineer.
Summary of Contents for NESTOMAT 6050
Page 31: ...GUIDO RAYOS X S A NESTOMAT 6050 31 MAN 012 Sept 00 Ed 1 Rev 3 1 2 2 3 4 5 6 7 8 9 10...
Page 37: ...GUIDO RAYOS X S A NESTOMAT 6050 37 MAN 012 Sept 00 Ed 1 Rev 3...
Page 38: ...GUIDO RAYOS X S A NESTOMAT 6050 38 MAN 012 Sept 00 Ed 1 Rev 3...
Page 47: ...GUIDO RAYOS X S A NESTOMAT 6050 47 MAN 012 Sept 00 Ed 1 Rev 3...