GUIDO RAYOS X, S.A.
NESTOMAT 6050
27
MAN-012
Sept.00
Ed. 1 / Rev. 3
suitable. If in doubt check with chemical supplier.
For chemical liquid immersion: Some chemicals can be harmful to plastics. If in doubt
check with chemical supplier.
There are no user serviceable parts inside this device.
Maintenance.
Only qualified personnel should carry out service and maintenance procedures.
After any maintenance is completed, ensure the equipment is functioning corectly in
accordance with the published performance specifications.
Disassembly
•
Remove the front panel, fixed by four screws.
•
remove gas inlet and outlet connectors using torque driver and special adaptor.
•
Unscrew two manifold retaining screws at bottom of unit.
•
Disconnet monometer tube at top ot T-connector by pushing down on collar of
T-connector and pulling tube free.
•
Remove manifold assembly.
Replacing valve assembly
•
Disassemble Neopuff as aoutlined above.
•
Remove the inspiratory valve assembly using the C-spanner. Valve unscrews
anti-clockwise.
•
Fit new inspiratory valve assembly, tighten clockwise with C-spanner. The
inspiratory valve assembly can be identified by the shite restrictor plug located
at the bottom end of the valve.
•
remove maximum pressure relief valve assembly using C-spanner. Unscrews
anti-clockwise.
•
Fit new maximum pressure relief valve assembly, tighten clockwise with C-
spanner. The masimum pressure relief valve has not restrictor plug.
•
Reassemble Neopuff and calibrate.
The manometer is not a serviceable item and must be replaced by Manometer kit
complete.
Summary of Contents for NESTOMAT 6050
Page 31: ...GUIDO RAYOS X S A NESTOMAT 6050 31 MAN 012 Sept 00 Ed 1 Rev 3 1 2 2 3 4 5 6 7 8 9 10...
Page 37: ...GUIDO RAYOS X S A NESTOMAT 6050 37 MAN 012 Sept 00 Ed 1 Rev 3...
Page 38: ...GUIDO RAYOS X S A NESTOMAT 6050 38 MAN 012 Sept 00 Ed 1 Rev 3...
Page 47: ...GUIDO RAYOS X S A NESTOMAT 6050 47 MAN 012 Sept 00 Ed 1 Rev 3...