GUIDO RAYOS X, S.A.
NESTOMAT 6050
24
MAN-012
Sept.00
Ed. 1 / Rev. 3
•
Please read and unsderstand the instructions fully before using the Neopuff
resuscitator and related accessories.
•
The Neopuff resuscitator is to be used only by persons trained in infant
resuscitation, It is responsibility of the purchaser to ensure that all users of this
device have been adequately trained in resuscitation techniques.
•
The Neopuff resuscitator should only be used after checking that correct
pressures will be delivered to the baby.
•
Ensure no smoking, naked flames or sources of ignition are present while the
unit is in use.
•
For connection to flow regulated oxygen or oxygen/air mixture only.
•
Minimum recommended operating gas flow 5 l/min. Do not attempt to use a
higher flow than 15 l/min.
•
The Maximum Pressure Relief can be adjusted up to a nominal 80 cm H
2
O, and
should only be done in exceptional circumstances by persons trained in infant
resuscitation. Do not attempt to set the Maximum Pressure Relief above 80 cm
H
2
O.
•
Use only a Fisher & Paykel T-piece.
1
Connect flow controlled oxygen or blender oxygen/air supply to the Neopuff
Gas Inlet port using oxygen inlet tubing.
2
Connect the patient supply tube to the Neopuff Gas Outlet port. If reusable
tubing is used, connect reusable T-piece to the patient, end of patient supply
tube.
3
Adjust the oxygen or oxygen/air flow between 5 and 15 l/min. Do not attempt to
use a higher flow that 15 l/min.
4
To check the Maximum Pressure Relief setting:
Conect the test lung to the T-piece, and check gas is flowing through the patient
Summary of Contents for NESTOMAT 6050
Page 31: ...GUIDO RAYOS X S A NESTOMAT 6050 31 MAN 012 Sept 00 Ed 1 Rev 3 1 2 2 3 4 5 6 7 8 9 10...
Page 37: ...GUIDO RAYOS X S A NESTOMAT 6050 37 MAN 012 Sept 00 Ed 1 Rev 3...
Page 38: ...GUIDO RAYOS X S A NESTOMAT 6050 38 MAN 012 Sept 00 Ed 1 Rev 3...
Page 47: ...GUIDO RAYOS X S A NESTOMAT 6050 47 MAN 012 Sept 00 Ed 1 Rev 3...