1 DISPL 16 CARS
2 DISPL 20 CARS
for LCD equipment press 1; for luminiscent display press 2.
II.10. ON/OFF SWITCH (SW) ACTIVATION
Press 0 key when applications programm is selected. The display will show:
WITH ON/OFF
WITHOUT ON/OFF
For equipments with ON/OFF switch on front panel keypad, press the ON/OFF key,
selecting WITH ON/OFF option.
II.11. CHANGE OF TEMPERATURE ALARM RANGE (HIGH /
LOW)
Press 4 key when applications programm is selected. The display will show:
RNG.TMP (0.1 ÷ 2.0)
R= __._ , H=
Preset value of temperature Range (R) (High / low) is 1 ºC. To this value corresponds a
hysteriris value (H) of 0,4 ºC.
To change the preset value (R) (between 0,1 ºC and 2,0 ºC), by means of the numerical
keypad, set the required value and press ENTER, the display will show the
corresponding hysteresis value (this value H can not be changed).
Then press ENTER again to accept the new values (R) and (H).