GUIDO RAYOS X, S.A.
NESTOMAT 6050
17
MAN-012
Sept.00
Ed. 1 / Rev. 3
1. SENSOR ALARM:
In case of air sensor failure or it gets disconnected or out of range; the
SENSOR Alarm will activate the corresponding LED and buzzer, the
alarm cannot be silenced. The LCD will inform to the attendant
displaying ,
* ALARM *
AIR SENSOR
The Radiant warmer is now out of service. Call to any authorized Service
engineer.
2. LOW TEMPERATURE ALARM:
Is activated when temperature reaches aproximately 1 ºC below the preset
temperature (See paragrap II.10 to change preset temperature alarm
range). "
LOW
" alarm LED will lit, to silence acoustic alarm press
SILENCE key. The LCD will display
* ALARM *
AT - RH OX
The alarm can be silenced by pressing the SILENCE key. If SILENCE
was pressed and the alarm conditions remain during a period of ten
minutes, the alarm will be activated again.
3. HIGH TEMPERATURE
Is activated when the temperature reaches aproximately 1 ºC higher than
the preset temperature (See paragrap II.10 to change preset temperature
alarm range). "
HIGH
" alarm LED will lit and the LCD will display ,
* ALARM *
AT + RH OX
to silence acoustic alarm press SILENCE key. If SILENCE was pressed
and the alarm conditions remain during a period of ten minutes, the alarm
will be activated again.
Check if heat sources are close to the Radiant warmer location, if positive
keep them away. If the alarm is still effective unplug the unit from the
power source, remove the patient from the radiant warmer and call to the
authorised service engineer.
B) SERVO CONTROL MODE
1. SENSOR ALARM:
Summary of Contents for NESTOMAT 6050
Page 31: ...GUIDO RAYOS X S A NESTOMAT 6050 31 MAN 012 Sept 00 Ed 1 Rev 3 1 2 2 3 4 5 6 7 8 9 10...
Page 37: ...GUIDO RAYOS X S A NESTOMAT 6050 37 MAN 012 Sept 00 Ed 1 Rev 3...
Page 38: ...GUIDO RAYOS X S A NESTOMAT 6050 38 MAN 012 Sept 00 Ed 1 Rev 3...
Page 47: ...GUIDO RAYOS X S A NESTOMAT 6050 47 MAN 012 Sept 00 Ed 1 Rev 3...