GUIDO RAYOS X, S.A.
NESTOMAT 6050
10
MAN-012
Sept.00
Ed. 1 / Rev. 3
II.OPERATION
IMPORTANT: B
EFORE TO USE THE
RADIANT WARMER
WITH A
PATIENT
,
IT IS NECCESSARY TO CHECK ITS GOOD PERFORMANCE CONDITIONS
,
AS WELL AS THOSE FROM THE ACCESSORIES AND OPTIONS
.
II.1 TO OPERATE
To start operation, plug the cord to the mains (check voltage) and press
the
I/O
switch located at the right side of the trolley.
Press the
ON/OFF
key at the front panel, the LCD will display
"GUIDO
RAYOS X - VER. x.x"
, immediately afterwards the Radiant warmer will
perform a selftest of all its circuitry.
II.2 TO PREHEAT
CAUTION:B
EFORE TO USE THE
RADIANT WARMER
WITH THE
NEWBORN
,
PREHEAT THE UNIT FOR
A BETTER COMFORT OF THE NEWBORN AND
AN ACCURATE PERFORMANCE
. P
REHEATING PERIOD OF TIME DEPENDS ON
ENVIRONMENTAL TEMPERATURE
. I
F PREHEATING MODE IS
NOT DESIRED
,
THE
OPERATION MODE
, A
IR OR
S
ERVO
,
CAN BE DIRECTLY SELECTED
.
Pressing any key the LCD will display,
AIR=__._ ºC
SELECT MODE
Press the key “
PRE
” to start the preheating until to reach aproximately
the factory preset temperature (28 ºC). In the meantime the LCD will
display
AT=__._ ºC (__._ ºC)
PREHEATING
After “
=
“ symbol we can find present temperature.
It is possible to modify preset preheating temperature in order to reach a
nearly temperature of the normal operation temperature. Proceed with
following steps to change preset preheating temperature:
•
Turn on the radiant warmer:
GUIDO RAYOS X
VER X.X
Summary of Contents for NESTOMAT 6050
Page 31: ...GUIDO RAYOS X S A NESTOMAT 6050 31 MAN 012 Sept 00 Ed 1 Rev 3 1 2 2 3 4 5 6 7 8 9 10...
Page 37: ...GUIDO RAYOS X S A NESTOMAT 6050 37 MAN 012 Sept 00 Ed 1 Rev 3...
Page 38: ...GUIDO RAYOS X S A NESTOMAT 6050 38 MAN 012 Sept 00 Ed 1 Rev 3...
Page 47: ...GUIDO RAYOS X S A NESTOMAT 6050 47 MAN 012 Sept 00 Ed 1 Rev 3...